Overview
HUITAI GDT-1000A high and low temperature alternating climate box is a professional temperature and humidity testing equipment, providing -20~ 150 ℃ Temperature range and 20%~ 98% relative humidity control. The equipment adopts a 7-inch TFT color LCD display, supports setting mode and program mode operation, and has the ability to set 100 sets of programs, 100 steps and 999 cycles.
Features
1. Adopt BTC equilibrate Temperature Control mode + DCC intelligent cooling capacity control + DEC intelligent electrical control
2. Equipped with a heated observation window with LED interior lighting
3. With RAM data logging function with battery protection, the maximum recording time is 350 days
4. Support RS-485 interface MODBUS RTU communication protocol and Ethernet interface TCP/IP communication protocol
5. Provide perfect safety protection system, including Compressor protection, humidification system protection and test chamber protection
6. Adopt French Taikang fully enclosed Compressor and high-efficiency refrigeration system
7. Equipped with high-efficiency SWEP plate type cold coal cold heat exchanger
Principle
The equipment adopts equilibrate Conditioning Control mode, through A-class PT100 armored Thermocouple for temperature and humidity sensing, and humidity control adopts wet-bulb temperature meter method. The refrigeration system adopts mechanical single-stage compression refrigeration or cascade Compressor refrigeration according to the model, and realizes precise temperature control through high-efficiency slope evaporator and SWEP plate cold and heat exchanger.
Applications
Electronic product testing, material research, automotive parts testing, aerospace parts testing, military product testing, scientific research experiments
Standards
GB/T5170.2-2008、GB/2423.1-2008、GB/2423.2-2008、GB/T5170.5-2008、GJB150.3、GJB150.4、GB2423.3-93、GB2423.4-93
Steps
1. Connect power 380V/50A
2. Set operating parameters through a 7-inch TFT color LCD display
3. Select setting mode or program mode to run
4. Temperature range -20~ 150 ℃ and Humidity range 20%~ 98%
5. Start the device running
6. Export data via USB or Ethernet interface
Notice
• The equipment needs to be used under the conditions of standard load capacity 1kg aluminum sheet and 200W heat generation
• Pay attention to Compressor overpressure protection, motor overheating protection and overcurrent protection
• Humidification system needs to prevent dry burning protection and abnormal water supply
• Regularly check the drying Filter status
• Ensure that the total power supply phase sequence and phase loss protection are normal
• Pay attention to the leakage protection and load short circuit protection of the whole machine
• The refrigeration pipeline adopts nitrogen flushing welding process, and leaks need to be checked regularly