Overview
DZG-6050 Vacuum oven is a factory direct sales desktop vacuum drying equipment, specially designed for powder drying and baking under vacuum conditions. It is especially suitable for processing heat sensitive, easy analysis, easy oxidation substances and complex components. It is widely used in industrial and mining enterprises, Universities & colleges, biological products, Food processing, agricultural research, chemical industry, ** units and various laboratories.
Features
1. The closure of the box door is elastic and adjustable, and the integrally formed silicone rubber door sealing ring ensures a high degree of vacuum in the box.
2, Using high performance CPU processing chip and High Sensitivity, High Accuracy Platinum Resistance Sensor temperature control system, temperature control precision, easy to operate.
3. Sensor fault alarm, over-temperature alarm, self-diagnostic dynamic control, temperature display correction, parameter memory and timing function for up to 9999 minutes.
4. Independent temperature limit control alarm system, automatically interrupts work beyond the limit temperature, providing double insurance.
Principle
Vacuum oven avoids thermosensitive material analysis or oxidation by reducing the boiling point in vacuum conditions to allow water or solvents in the material to evaporate at a lower temperature. It uses four-week heating, combined with a High Accuracy Platinum Resistance Sensor and a microprocessor with fuzzy PID control technology to achieve precise temperature control.
Applications
Industrial and mining enterprises, Universities & colleges, Biological products, Food processing, Agricultural research, Chemical industry, ** units, various laboratories
Steps
1. Adjust the closure and tightness of the box door to ensure the sealing of the silicone rubber door sealing ring.
2. Set parameters such as temperature and time, and operate with a digital display and push-button controller.
3. Start the equipment and monitor temperature fluctuations and vacuum levels.
4. During the operation of the equipment, if there is an overtemperature or Sensor failure, the system will automatically alarm and interrupt the work.
5. After drying, turn off the equipment and remove the material.
Notice
• Ensure that the working environment temperature is within the range of 5~ 30 ° C, relative humidity ≤ 80%, and altitude ≤ 1500m.
• Use AC220V + 10%, 50Hz +/- 1Hz power supply to avoid voltage instability.
• Pay attention to over-temperature alarm and Sensor fault prompt during operation, and deal with abnormalities in time.
• Regularly check the sealing and vacuum of the box door to ensure the normal operation of the equipment.
• Avoid opening the box door while the equipment is running to prevent vacuum failure and safety risks.