Overview
DHG-401A aging test chamber is a factory direct product, mainly used in industrial and mining enterprises, laboratories and R & D institutions for aging, drying, baking, wax, sterilizing and other operations.
Features
1. The temperature control system with high performance CPU processing chip and High Sensitivity and High Accuracy Platinum Resistance Sensor makes the temperature control more accurate and the operation more convenient.
2. Sensor fault alarm, over-temperature alarm, self-diagnostic dynamic control, temperature display correction, parameter memory and timing function for up to 9999 minutes.
3, the use of high-quality Cold rolled steel plate to make the shell, using phosphating electrostatic spraying technology, more durable, more convenient to clean.
4. Double-decked tempered Glass observation window that meets CCC certification is safer and clearer for samples in the observation box.
Principle
Precise temperature control is realized through high performance CPU processing chip and High Accuracy Platinum Resistance Sensor, forced circulation convection air duct is used to ensure uniform temperature distribution, and fuzzy PID control technology is supported for microprocessor temperature management.
Applications
Industrial and mining enterprises, laboratories, R & D institutions
Standards
CCC certification
Steps
1. Connect the power supply to AC220V +/- 10%, 50Hz +/- 1Hz.
2. Set the temperature control range to Rt + 10 to 300 ° C.
3. Optional timing function, range from 0 to 9999 minutes.
4. Operate through the LCD display and push-button interface.
Start the device, monitor Temperature Fluctuation +/- 5 ℃ and alarm setting SV + 5 ℃.
Notice
The working environment temperature should be kept at 5~ 30 ℃.
• The relative humidity should not exceed 80%.
• The altitude does not exceed 2000m.
• Power supply voltage AC220V +/- 10%, 50Hz +/- 1Hz.
• Note that the power consumption of the device is 1500W.
• Regularly check whether the Sensor and Alarm functions are normal.
• Use appropriate methods when cleaning to avoid damage to the shell coating.