Overview
MS102 Halogen moisture meter is a fast moisture measurement instrument with color TFT large-screen Touchscreen. It can be widely used in the production and experimental process of plastics, chemical industry, food, grain, feed, seed, Tobacco, tea, Spinning & weaving, agriculture and forestry, papermaking, rubber and other industries for rapid determination of sample moisture content.
Features
1. Adopt color TFT large-screen Touchscreen, which displays rich information and is easy to read.
2. With weight, temperature, clock calibrating function, provide manual, timing, automatic three measurement modes, heating end buzzer prompt.
3. Using a high-mass halogen lamp as a heating source, the temperature rises rapidly, the heating is uniform, and the sample test is fast. Generally, the sample can be measured in a few minutes.
4. High Accuracy Thermal Insulation Sensor is used to collect data and temperature reliably, effectively ensuring the authenticity and accuracy of sample data.
5. Built-in RS232 output interface, can be connected to computers, printers and other external devices.
Principle
The instrument uses the principle of thermogravimetry. The built-in halogen lamp heats the sample quickly and evenly, causing the water in the sample to evaporate. At the same time, the High Accuracy Sensor continuously monitors the weight change of the sample during the heating process. By calculating the weight difference of the sample before and after heating, the instrument can quickly and accurately calculate the moisture content percentage of the sample.
Applications
Plastics, chemicals, food (surimi, dehydrated vegetables, meat and aquatic processing, noodles, flour, biscuits, mooncakes, etc.), grain, feed, seeds, rapeseed, Tobacco, tea, Spinning & weaving, agriculture and forestry, paper, rubber
Standards
Not mentioned in the data
Steps
1. Start the machine to preheat, and set the appropriate heating temperature (adjustable from 60 ° C to 200 ° C) and heating time (adjustable from 0 to 99 minutes) according to the sample type.
2. Put the sample into the sample tray, and the instrument automatically records the initial weight.
3. Select the measurement mode (manual, timed or automatic) and start the heating.
4. After heating, the buzzer prompts, and the instrument screen displays the measurement results, including moisture content percentage, dry weight content percentage or dry matter weight.
Notice
The instrument should be placed on a stable, vibration-free surface for use.
The sample weight is recommended to be not less than 2 grams to ensure measurement accuracy.
The heating temperature setting should be reasonably selected according to the characteristics of the sample to avoid coking the sample due to excessive temperature.
Regularly calibrate the weight and temperature of the instrument to ensure the accuracy of the measurement data.