Overview
HI97771 residual chlorine [free chlorine] -UHR total chlorine Portable Photometer meets IP67 waterproof standards and is designed with an innovative Optical inspection system for good accuracy and repeatability. The instrument has CAL CheckTM function, allowing performance verification and calibrating using NIST traceability standards, built-in reaction Timer to ensure measurement conformity, GLP function to display calibrated date and time, and equipped with battery status indicator and automatic shutdown settings.
Features
1. Compliant with IP67 waterproof standard
2. Optical inspection system with high accuracy and repeatability
3. CAL CheckTM function supports NIST traceability standard verification and calibrating
4. Built-in Reaction Timer ensures measurement conformity
5. GLP function shows the last calibrated date and time
6. Equipped with battery status indicator and automatic shutdown function
7. Automatically store the last 50 measurements
8. Battery life up to 800 hours (no backlight)
Principle
The instrument is based on the EPA 330.5 DPD method standard, using Light Emitting Diode @525nm Illuminant for measurement, Bandwidth 8nm, wavelength accuracy +/- 1.0nm. Absorbance of the sample at a specific wavelength is detected by an Optical inspection system, and residual chlorine and total chlorine concentration values are calculated.
Applications
Water quality testing, environment monitoring, drinking water treatment, swimming pool water treatment, industrial water treatment
Standards
IP67 waterproof standard, EPA 330.5 DPD method standard
Steps
1. Prepare the sample and load it into the reaction cup
2. Turn on the instrument and select Measurement mode
3. Use CAL CheckTM to calibrate verification
4. Start measurement, the instrument automatically timed the reaction
5. Read and record measurement results
6. Instrument automatically stores measurement data
Notice
• Make sure the instrument is calibrated before use
• The reaction cup needs to be kept clean to avoid contamination
• Ensure that the sample temperature is within the range of 0-50 ° C during measurement
• Avoid use in environments where Humidity exceeds 100% RH
• If you don't use it for a long time, please take out the battery in time.
• Use only 1.5V AA batteries of specified specifications
• Automatic shutdown function can be started after 15 minutes of no operation to save power