Overview
The LANYU LYD500W-T Touch Homogenizer (110V) is a device for laboratory Sample Handling. It realizes mixing, uniform dispersion, emulsification and pulverization of samples through high-speed motion, and is suitable for a variety of application scenarios from cell crushing to food preparation.
Features
1. Equipped with touch knob and TFT-LCD display, it can display operating parameters such as speed, timing, torque Linear dispersion, voltage, current, frequency and power in real time.
2. Provide 12G, 18G, 24G, 30G, 36G various specifications of SS316 stainless steel dispersion emulsification work head, processing capacity from 30ml to 13000ml, adapt to different viscosity and volume of samples.
3. Adopt stepless speed regulation, with a wide speed range of 1000 to 28000rpm, and the maximum linear speed of the rotor can reach 26m/s.
4. With Timer Range 1~ 1000 minutes.
5. Built-in real-time overload protection function ensures the safe and stable operation of the equipment.
6. The working head bearing is made of PTFE material, allowing the operating temperature to be up to 120 ° C.
Principle
The Homogenizer uses a motor to drive the working head to swirl/spin at high speed, generating strong shearing, Impact and cavitation, so that the particles or droplets in the sample are crushed, dispersed and uniformly mixed, thereby achieving emulsification, homogenization or cell crushing and other effects.
Applications
Cytological studies (cell crushing, cell dissolving, extraction of biomolecules), Food processing (dairy products, sauces, jam preparations), solution mixing, particle dispersion, chemical reactions
Standards
Duty Cycle meets S1 (continuous) standard
Steps
1. Select the appropriate dispersion emulsification head according to the sample volume and viscosity and install it.
2. Ensure that the liquid height in the sample container is between the minimum and maximum immersion heights specified by the working head.
3. Connect the AC110V power supply.
4. Set the required speed (1000~ 28000rpm) and timing time (1~ 1000min) through the touch knob.
5. Start the device and monitor the operating parameters through the TFT-LCD display.
6. After running, turn off the power supply, remove and clean the working head.
Notice
* Please select the working head of the corresponding specification according to the sample characteristics (volume, viscosity).
When operating, be sure to ensure that the liquid immersion height is within the specified range (e.g. 35mm~ 150mm for 12G) to avoid idling or excessive immersion.
* Ambient Temperature should be ≤ 40 ℃, relative humidity should be ≤ 80% RH.
Equipment is protected against overload, but do not overload for long periods of time.
When cleaning and maintaining, make sure the equipment is powered off.
The working head and the parts in contact with the sample are made of stainless steel. Please use the appropriate cleaning method.