Overview
HUXI HLS-1312F High Accuracy Chiller is a high-end model designed for high-quality analytical instrument needs, mainly used in atomic absorption instrument, Light spectrum instrument, mass spectrometry, SEM, lens electron microscope, Soxhlet extraction instrument, diffusion pump, molecular pump, Kjeldahl nitrogen meter, ion thinning instrument, Fourier transform mass spectrometry (FIS), magnetron sputtering instrument, small vacuum coating machine and other analytical instruments.
Features
1. High-performance circulating water pump, the pressure of the circulating pump can be adjusted steplessly
2. Removable dust net cover, easy maintenance, easy user operation, quick installation and debugging
3. Refrigeration function and circulation function can be controlled independently or simultaneously
4. The combination of proportional refrigeration technology, hot gas bypass technology and PID control technology has high temperature control accuracy
5. With multiple Safety protection such as temperature protection, water pressure protection, overheating and overcurrent protection to ensure experimental safety
6. Multiple operation modes are available, accidental power failure can be automatically restored to ensure continuous supply of chilled down circulating water
7. Using imported Compressor and ultra-quiet fan, the whole machine has low working noise and is suitable for laboratory environment
8. Alarm reminder function, light and sound reminder at the same time, the code directly displays the error point
9. Multi-purification water quality design: water tank infiltrating device, waterway material selection, standard water Filter purification
Principle
The chiller achieves precise temperature control through the combination of proportional refrigeration technology, hot air bypass technology and PID control technology, and uses a high-performance circulating water pump to provide an adjustable pressure chilled down water cycle to ensure the safe operation of the equipment through multiple protection mechanisms.
Applications
Atomic Absorption Instrument, Light Spectrometer, Mass Spectrometry, SEM, Lens Electron Microscope, Soxhlet Extraction Instrument, Diffusion Pump, Molecular Pump, Kjeldahl Nitrometer, Ion Thinning Instrument, Fourier Transform Mass Spectrometry (FIS), Magnetron Sputtering Instrument, Small Vacuum Coating Machine
Steps
1. Install the equipment and connect the inlet and outlet pipes
2. Adjust the circulating pump pressure as required
3. Select the control mode for refrigeration and circulation functions
4. Set the required Temperature range (5~ 35 ℃)
5. Start the device running
6. Monitor the temperature parameters on the LCD screen
7. Regularly clean the removable dust net cover
Notice
• Regularly check and clean the dust net cover
• Pay attention to the alarm prompt and deal with the fault in time
• Ensure that the water quality meets the requirements and regularly check the filtrating device
• Avoid equipment operating under conditions exceeding the rated parameters
• Keep equipment well ventilated