Overview
HUXI HLC-2027SH High and Low Temperature Circulating Bath is a vertical equipment widely used in universities and enterprises in the fields of biology, chemical industry, chemistry, life science, physical property testing and other industries for Quality control. It provides a heat-cooled controlled, uniform and constant temperature field source for user experiments and production. It can be used to perform constant temperature tests or tests on specimens or products, and can also be used as a direct or auxiliary heating and cooling heat source or cold source.
Features
1. Maintenance-free pressure/suction pump, resistant to fiber and metal particulate matter, can be stirred in both horizontal and vertical directions to make the mixing more uniform;
2. Multiple safety protection: low liquid level protection, which can be alerted by light and sound at the same time when alarming;
3. Parts in contact with liquids or vapors are made of high-performance stainless steel and high-performance plastics.
4. Pt100 temperature sensor, single point and two point temperature calibrating;
5. Can be domestic circulation, domestic circulation standard with two different heights of the sample cup bracket, suitable for different sample cup in the bath thermostatic and viscosity test;
6. Can be external circulation, to achieve the Viscometer water bath jacket for thermostatic, suitable for Viscometer in the bath outside the bath jacket sample viscosity test;
7. The buzzer can be set to silent or audible.
Principle
The High and Low Temperature Circulating Bath is temperature controlled by the built-in heating and cooling system (using R134A refrigerant), and the temperature sensor (Pt100) monitors and calibrates the temperature to ensure that the temperature fluctuation value is within +/- 0.03 ° C. Through the internal and external circulation system, the pumped bath is evenly stirred and temperature distributed, providing a stable source of hot and cold field.
Applications
Biological, chemical, chemical, life sciences, physical property testing
Steps
Connect the power supply to ensure that the voltage is within the range of 200-230V.
2. Check the liquid level to ensure that it is not lower than the safety standard.
3. Set the required temperature through Touchscreen with an accuracy of +/- 0.1 ° C;
4. Start the equipment and choose the domestic circulation or external circulation mode;
5. Place the sample cup or connect the Viscometer jacket;
6. Monitor the temperature display to ensure stability at the set value.
7. After the test is completed, turn off the device and disconnect the power.
Notice
- Operating Environment Temperature should be between 5 and 32 ℃;
- Pay attention to low liquid level protection to avoid dry burning of equipment.
- Regularly check the circulatory system to prevent blockages.
Avoid running for long periods of time under high or overheated conditions.
- Device weight, Net weight 43KG, gross weight 51KG;
- The buzzer settings can be adjusted to the sound or silent mode as required.