Please contact customer service
Explosion-proof box and integral explosion-proof electric device, Explosion-proof level up to ExdembIIBT4Gb, equipped with explosion-proof heating pipe and flexible connecting pipe, effectively isolating electrostatic sparks; Studio volume 230L, Temperature uniformity up to +/- 2.5%, to ensure safe and stable operation in high temperature environment.
Category : Drying Oven
Brand : TaiSiTe
SKU : NB047635
Service :The delivery and service By NBCLab.Com
TAISITE TEX-230LG explosion-proof blast drying oven is a laboratory equipment specially designed for environments with explosion-proof requirements. It is mainly composed of drying oven body, explosion-proof heating system and explosion-proof electrical control parts. The box is designed with explosion-proof type, and the electrical appliance adopts the overall explosion-proof technology, which can effectively prevent the leakage of unsafe factors such as high temperature, static electricity, and sparks to ensure safe operation.
1. The box is made of explosion-proof type, the equipment and electrical appliances use the overall explosion-proof technology, and the explosion-proof control box is centrally protected to prevent the leakage of high temperature, static electricity, sparks and other unsafe factors. 2. The inner tank of the box body is made of mirror stainless steel, and the shell is made of high-quality steel plate, which is beautiful and generous in shape, and the liner is easy to clean. 3. Vertical double air duct air jacket, turbine centrifugal forced air supply, uniform temperature. 4. Pull-out tray Shelf, sampling is convenient and fast. 5. LCD multi-data full-screen operation display, with power failure memory function. 6. Convenient operation, support fixed-value operation, regular operation, automatic stop, and appointment start. 7. Use explosion-proof 304 stainless steel heating pipe heating. 8. Exposed connecting wires are all explosion-proof flexible connecting pipe sets to isolate static sparks and temperature leakage to ensure safety. 9. With over-rise alarm, menu lock, over-rise prevention function.
The equipment uses explosion-proof heating pipes to heat the air in the box, and uses vertical double air ducts and turbine centrifugal fans to force air supply to form uniform hot air circulation, so that the temperature in the working room quickly reaches the set value and remains stable. The explosion-proof electrical Control system precisely controls the temperature and operating procedures to ensure safe and stable operation in flammable and explosive environments.
Petroleum, chemical, medical, scientific research, laboratory
GB/T30435-2013
1. Turn on the power supply (AC220V). 2. Set the required temperature, time and other parameters through the LCD screen and explosion-proof four buttons. 3. Select the operation mode (e.g. fixed-value operation, regular operation, etc.). 4. Start the equipment, start heating and blast operation. 5. After the operation is completed, the device can be automatically stopped. 6. After use, turn off the power.
* Equipment should be installed in a place that meets explosion-proof requirements and ensure that it is well ventilated. * Power supply voltage and equipment Rated voltage (AC220V) before use. * Do not overload, single shelf load-bearing no more than 15kg. * When the equipment is running, do not switch the door frequently, so as not to affect the Temperature uniformity and safety. * When cleaning the liner, please use a soft cloth to wipe and avoid the use of corrosion cleaners. Non-professional personnel should not disassemble or modify the explosion-proof components and electrical circuits of the equipment. If there is a malfunction alarm (such as an over-lift alarm), it should be stopped immediately and a professional should be contacted for maintenance. * When not in use for a long time, the power supply should be cut off and the equipment should be placed in a drying environment.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.