Using the principle of infrared NDIR, the detection accuracy is +/- 3% F. S, and the Response Time is less than 25 seconds. The system integrates a pretreatment unit, which can cope with severe working conditions such as moisture and dust. It supports 4-20mA, RS485 output and wireless module expansion to realize real-time data upload and equipment linkage.
Category : Gas Detector
Brand : Skyeaglee
SKU : NB047911
Service :The delivery and service By NBCLab.Com
SK-7500-LEL-JY LEL on-line monitoring system is an on-line monitoring system for continuous monitoring of LEL (lower explosive limit) content of organic matter in pipelines. The system integrates gas sampling, preprocessing, detection and data management. It has a simple structure, wide dynamic range and strong real-time performance. It can realize on-site unattended and can communicate with the enterprise DCS system.
1. Pre-processing, High Accuracy, Small Overall Size, Easy Installation. 2. Professional gas pretreatment for humid, tar, dust and other harsh environments. 3. Using the original high-performance Sensor, long life, good anti-interference performance, accurate and stable monitoring data. 4. Equipped with an infrared remote control, full-function operation can be achieved without opening the lid. Standard 4-20mA, RS485 and a set of relay output signals, easy to access industrial control systems such as PLC, DCS or stand-alone control. 6. Optional wireless modules such as DTU and LORA can be used to achieve real-time data upload. 7. Optional 7-inch industrial touch screen, intuitive and clear display of real-time concentration and Linear dispersion map.
The system adopts active sampling method. After the pipeline gas is extracted by the sampling pump, it is pretreated by precise filtrating, cooling and dehumidification, flow control, etc., to obtain clean, drying and constant sample gas, and then sent to the LEL concentration detector. The detector is based on the principle of infrared NDIR (non-spectral infrared), using the absorption characteristics of different gas molecules to specific wavelengths of infrared light, according to Lambert-Beer's law, the gas components are identified and their LEL concentration is determined by measurement of Absorption intensity.
Generally used for RTO front-end pipeline monitoring to prevent explosion hazards due to high concentration
GB3836.1-2010、GB3836.2-2010、GB 3836.15-2000、GBT50493-2019、GB12358-2006、GBZ 2.1-2007、GB 4208-2008、GB 16297-1996、GB/T 16157-1996、GB/T 5468-1991、HJ/T 397-2007、HJ/T 76-2017、HG/T 20507-2014
1. Determine and install the sampling Probe according to the selection point requirements (e.g. avoid elbows, valves, select vertical pipe sections, etc.). 2. Connect the system power supply (AC220V) and signal output cable (such as 4-20mA, RS485). 3. If the Back flushing function is optional, connect the clean instrument air source (pressure 0.4~ 0.7MPa). 4. The system is powered on, the sampling pump is started, and the sample gas enters the detector after pretreatment. 5. View the real-time concentration through the display screen or infrared remote control, and perform operations such as parameter setting and calibration. 6. The system data can be connected to the industrial control system through the output signal, or uploaded to the server through the optional wireless module.
• Applicable pipeline working conditions are: temperature ≤ 80 ° C, no or trace dust and water vapor, micro negative pressure, normal pressure or positive pressure, no explosion-proof requirements. If the working conditions exceed, non-standard design is required. • Installation of sampling points should prioritize the vertical pipe section and the negative pressure area of the flue to ensure sample representativeness. • The measurement position should avoid the flue elbow and the rapidly changing part of the section. Please refer to the materials for specific distance requirements. • The system power supply is AC220V, please pay attention to safety when wiring. • It is recommended to use Shielding cable for signal transmission cable. For long-distance transmission, the wire diameter should be thickened and the Shielding layer grounded. • If the Back flushing function is selected, the air source needs to be clean, oil-free and water-free instrument wind. Equipment requires vibration, and the installation environment should meet ≤ 100 Hz (0.3m/s). • It can be operated by infrared remote control during maintenance to avoid unnecessary opening of the lid. • Sensors have a lifespan of 2 to 6 years, depending on the usage, and require regular maintenance and calibrating.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.