$ 409.00Ex works
MCU chip to achieve low power consumption operation, equipped with intelligent Sensor to ensure detection accuracy, Response Time ≤ 30 seconds, with flameproof structure to adapt to dangerous environment, support a variety of signal output for System integration.
Category : Gas Detector
Brand : GUOYUGR
SKU : NB048770
Service :The delivery and service By NBCLab.Com
Ethanol detector GY/VGD-200-C2H6O-G (0-50ppm) is a low-power Modular design using MCU chip, with intelligent Sensor detection technology, overall flameproof (d) structure, fixed Installation method detector. Standard configuration is LCDTFT high definition LCD display, three-wire system 4~ 20mA analog, RS485 digital and switching signal output, provide customized products according to user requests, and also support Output signal fine-tuning and other functions, convenient system networking and maintenance.
1. Customizable Range Intelligent Sensor is adopted to ensure the advantages of high Sensitivity, high accuracy, linear stability and fast Response of the detector 2. The MCU chip is used, which highlights the characteristics of low power consumption, safe and reliable performance, accurate data conversion, and stable signal output of the detector 3.4-20mA, RS485, switch and other signal output, easy access to PLC, DCS, DDC and other industrial control systems 4. Support customized wireless signal transmission: Zigbee, LORA, 433M, 4G, WiFi, NB, GPRS, etc 5 point array high definition LCDTFT LCD screen, Chinese and English Operating interface can be switched 6. The alarm value, zero adjustment, calibrate, reset, modify the address code and other function menu operations of the detector can be performed by remote control without opening the lid 7. Local integrated sound and light alarm indication, remote mobile phones, computers, monitoring rooms, third-party platforms and other alarm prompts can be realized through alarm control host or wireless signal 8. Aluminum alloy cast explosion-proof shell, sturdy and durable 9. Passed a variety of CMA, CNAS, CE, ROHS, SIL, UL, ECM safety and other international certifications
The ethanol detector adopts the principle of catalytic combustion detection, and detects the concentration of C2H6O gas in the environment by diffusion Sampling method. The detector converts the detected gas concentration signal into standard signal output through MCU chip processing.
Industrial Gas detection, environment monitoring, safety protection, chemical production, Petrochemicals, coal mine safety, dust environment, dangerous places
GB/T3836.1-2021、GB/T3836.31-2021、GB3836.15-2000、GB12476.1-2013、GB12358-2006、GBZ2.1-2007、GB4208-2008
1. Install the detector to the designated location 2. Connect the power cord and work cable 3. Set the alarm value through the remote control 4. Zero and calibrate operations 5. Output signal parameters 6. Start the detector and start working 7. Monitor detection data through display screens or remote devices
• Ensure that the electrical interface is well sealed during installation • Ambient Temperature should be kept in the range of -20~ 50 ℃ • Environmental humidity should be controlled at 10-95% RH (no condensation) • When the cable transmission distance exceeds 1000 meters, the single wire diameter must be ≥ 1.5mm. • Shielding must be connected to the earth Sensors have a lifespan of 2 to 3 years and need to be replaced regularly Explosion-proof level of Ex db IIC T6 Gb, to be used in locations that meet explosion-proof requirements • Alarm settings can choose regular high and low alarms or interval control alarms as needed
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.