Touchscreen thermostatic liquid Density Meter JHY-1000 is an instrument that combines Peltier temperature control technology to control the temperature of the sample liquid at 20 ° C, and determines the Density by electromagnetically triggering U-shaped tube oscillation and measuring the frequency. It can test most liquids (except high viscosity and liquids that corrode Glass such as hydrogen fluoride). It has the advantages of practicality, reliability, high accuracy and high measurement accuracy. It is widely used in beverage and food, Petrochemicals, Inspection & Quarantine, Measurement calibrate and other fields. The Density Meter also supports Density concentration conversion software to directly read concentration values, and integrates functions such as calibrate, sampling, thermostatic control, testing, cleaning, drying, operation prompts, observation window, Density Linear dispersion, sample library, data storage, data statistics, and result printing, making it easier to use and maintain.
Features
1. Touchscreen: Direct touch screen operation, convenient setting and rich functions.
2. Thermostatic control: Combined with Peltier temperature control technology, the test is more accurate.
3. Peristaltic pump: no leakage backflow, constant flow, high and stable repeability test accuracy.
4. Sample library: 100 sets of sample preset attributes can be stored for test selection.
5 high definition real-time video: easy to observe the U-shaped glass tube test, sample, cleaning and drying.
6 calibrating function: use pure water as a standard calibrating instrument to deal with air value drift.
7. Density Linear dispersion: Real-time observation of sample Density changes.
8. Data storage: 100 sets of test records can be stored, and the average and maximum MIN Minimum can be obtained by counting the most recent n sets of records.
9. Results printing: Equipped with a printer, test results can be printed.
10. Miscellaneous: Optional anti-corrosive hose to meet corrosive liquid testing requirements.
Principle
The thermostatic electronic liquid Density Meter controls the sample liquid temperature at 20 ° C by combining the Peltier temperature control technology, and then the U-shaped tube is vibrated by electromagnetism, and the Density of the sample liquid is calculated by the oscillation frequency.
Applications
Beverages, Petrochemicals, Inspection & Quarantine, Measurement calibrating
Standards
Drying air/distillation water
Steps
1. calibrate: use pure water as the standard for instrument calibrating.
2. sampling: by peristaltic pump sample, the capacity is less than 6ml.
3 thermostatic control: set the temperature range 15~ 35 ℃ to ensure accurate testing.
4. Test: Select the measurement mode (such as Density g/cm3, concentration%, Baume, etc.) for measurement.
5. Cleaning: Clean the U-shaped glass tube.
6 drying: drying sample tube.
7. Data storage: Save test records, can store 100 sets.
8. Results printing: Use the printer to output the test results.
Notice
Avoid testing high viscosity liquids or liquids such as hydrogen fluoride that can corrode Glass.
• Regularly use pure water calibrating instruments to prevent air value drift.
• When testing corrosive liquids, choose anti-corrosive hoses.
• Observe the U-shaped glass tube through high definition real-time video during operation.
• Power frequency range is 110V~ 240V, 50/60Hz, power 100W.