Overview
Electronic Balance is a High Accuracy weighing device with Capacity 220g and readability accuracy of 0.1mg. It adopts a new generation of electromagnetic force Sensor, equipped with a 5-inch color TFT touch display, supports external calibrate and a variety of weighing units to meet the accurate weighing needs of the laboratory.
Features
1. New Generation High Accuracy, High Resolution Electromagnetic Force Sensor, Providing Stable and Reliable Weighing Results
2.5-Inch color TFT touch display, supports operation and use with laboratory gloves
3. Fully automatic internal calibrating for external calibrating or time and temperature triggering (AE-C series)
4. Chinese and English interface switching, front horizontal bubble adjustment, simple and intuitive operation
5. Sensor transport protection device to ensure its safety
6. Various weighing units: mg/g/ct/oz, etc
7. Optional lower hook design to meet users' special application needs
8. Standard RS232 interface, optional PC-STWC function to transfer weighing data directly to open applications such as Excel
9. Built-in date and time display to meet traceability and compliance requirements
10. Power supply: 110V-240V, 50Hz/60Hz
Principle
Using electromagnetic force Sensor technology, accurate weighing is achieved through the principle of electromagnetic force equilibrating. When the item is placed on the pan, the Sensor generates the corresponding electromagnetic force to equilibrate the item gravity, and calculates the item mass by measuring the electromagnetic force.
Applications
Laboratories, scientific research institutions, pharmaceutical, chemical, food testing, jewelry weighing, educational institutions
Steps
1. Place the Balance on a stable countertop and adjust it to a horizontal state through the front horizontal bubble
2. Connect the Power Adapter and turn on the power switch
3. Wait for the Balance self-test to complete, and the display screen shows zero.
4. Select the desired weighing unit (mg/g/ct/oz, etc.)
5. Place the item lightly in the center of the pan
6. Read the weighing result after waiting for the value to stabilize
7. If you need to peel, press the peel button and weigh again.
8. Turn off the power after use
Notice
• Please place the Balance on a stable, vibration-free countertop for use
• Be sure to adjust the horizontal bubble to a horizontal state before use
• Avoid use in strong electromagnetic interference environments
• Capacity must not exceed 220g
• Keep the pan clean and avoid contact with corrosive substances
• Ensure Sensor Protection is locked during transportation
• Ambient Temperature should be kept within 10~ 30 ℃
• Regularly calibrated to ensure weighing accuracy