Overview
FA 20.04 million Electronic Balance is a High Accuracy weighing device from the Seiji brand, featuring a new generation of electromagnetic force Sensor technology with readability accuracy of 0.1mg and 200g Capacity. The product features an oversized backlit LCD display and supports a variety of weighing units to meet the precise weighing needs of laboratory and industrial environments.
Features
1. New Generation High Accuracy, High Resolution Electromagnetic Force Sensor, Providing Stable and Reliable Weighing Results
2. Large backlit high definition LCD LCD display, weighing results are clearly displayed
3. Dynamic temperature compensation technology to ensure the accuracy of weighing
4. External calibrating function
5. The weighing stability time and reaction speed are adjustable, and the reaction time is < 3S.
6. Various weighing units: mg/g/ct/oz, etc
7. Optional lower hook design to meet users' special application needs
8. Standard RS232 interface, optional PC-STWC function to transfer weighing data directly to open applications such as Excel
9. Built-in date and time display to meet traceability and compliance requirements
10. Power supply: 110V-240V, 50Hz/60Hz
Principle
FA2004 Electronic Balance uses electromagnetic force Sensor technology to achieve accurate weighing through the principle of electromagnetic force equilibrating. When the item is placed on the pan, the Sensor generates the corresponding electromagnetic force to equilibrate the gravity of the item, and uses dynamic temperature compensation technology to ensure the accuracy of weighing results at different Ambient Temperatures.
Applications
Laboratory, industrial weighing, quality control, R & D testing
Standards
Meet traceability and compliance requirements
Steps
1. Place Balance on a Stable Workbench
2. Connect Power Adapter (input 110-240V, output 12V 1A)
3. Turn on and warm up to a stable state
4. External calibrating as required
5. Place items to be weighed on the pan
6. Read the weighing results on the LCD display
7. Data can be transferred to the computer through the RS232 interface
Notice
• Ambient Temperature should be kept within 10~ 30 ℃
• Ensure Balance is placed on a smooth and vibration-free countertop
• Preheat and calibrate before use
• Avoid use in strong electromagnetic interference environments
Regular cleaning of pan and Draft Shield
• Power Adapter input voltage range is 110-240V