Overview
The AE323C Electronic Balance is a High Accuracy weighing device with a Capacity of 320g and readability accuracy of 1mg. It adopts a new generation of electromagnetic force Sensor to ensure stable and reliable weighing results. Equipped with a 5-inch color TFT touch display, supports glove operation, and has built-in auto calibration function to meet the needs of laboratory and industrial applications.
Features
1. New Generation High Accuracy, High Resolution Electromagnetic Force Sensor, Providing Stable and Reliable Weighing Results
2.5-Inch color TFT touch display, supports operation and use with laboratory gloves
3. Externally calibrated or fully automatic internal calibrating triggered by time and temperature
4. Chinese and English interface switching, front horizontal bubble adjustment, simple and intuitive operation
5. Sensor transport protection device to ensure its safety
6. Various weighing units: mg/g/ct/oz, etc
7. Optional lower hook design to meet users' special application needs
8. Standard RS232 interface, optional PC-STWC function to transfer weighing data directly to open applications such as Excel
9. Built-in date and time display to meet traceability and compliance requirements
10. Power supply: 110V-240V, 50Hz/60Hz
Principle
The Electronic Balance uses electromagnetic force Sensor technology to achieve accurate weighing through the principle of electromagnetic force equilibrating. When an item is placed on a pan, the Sensor detects weight changes and converts them into electrical signals, which are processed internally and displayed on the display. Auto calibration is triggered by time or temperature to ensure long-term accuracy.
Applications
Laboratory, industrial weighing, quality control, R & D, education, pharmaceutical, chemical, food, jewelry weighing
Steps
1. Place the Balance on a smooth countertop and adjust the level through the front horizontal bubble
2. Connect Power Adapter, input voltage 110-240V
3. Select the Chinese and English interface after booting, and set the unit as needed
4. Automatic or external calibrating to ensure accuracy
5. Place the sample in the pan for weighing, and use the peeling function to deduct the container weight
6. Transfer data to external devices through RS232 interface or PC-STWC function
Notice
• The operation temperature range is 10~ 30 ℃, avoid using in extreme environments
• Ensure that the Power Adapter input voltage matches the requirements of 110-240V
• Check the horizontal bubble before use and keep the Balance level
Periodically calibrated to maintain accuracy
• Sensor protection to prevent damage during transportation
• Avoid overloading, Capacity does not exceed 320g
• Use a soft cloth when cleaning to avoid liquids from entering the equipment