Overview
Shangtian Jingyi JA5003C One thousandth Electronic Balance is a High Accuracy weighing equipment with Capacity 500g and readability accuracy 1mg. It adopts a new generation of electromagnetic force Sensor and dynamic temperature compensation technology to ensure stable and reliable weighing, suitable for laboratories, industries and other scenarios.
Features
1. New Generation High Accuracy, High Resolution Electromagnetic Force Sensor, Providing Stable and Reliable Weighing Results
2. Large backlit high definition LCD LCD display, weighing results are clearly displayed
3. Dynamic temperature compensation technology to ensure the accuracy of weighing
4. Externally calibrated or fully automatic internal calibrating triggered by time and temperature
5. The weighing stability time and reaction speed are adjustable, and the reaction time is < 3S.
6. Various weighing units: mg/g/ct/oz, etc
7. Optional lower hook design to meet users' special application needs
8. Standard RS232 interface, optional PC-STWC function to transfer weighing data directly to open applications such as Excel
9. Built-in date and time display to meet traceability and compliance requirements
10. Power supply: 110V-240V, 50Hz/60Hz
Principle
The Electronic Balance is based on the Electromagnetic Force Sensor technology and achieves accurate weighing through the principle of electromagnetic force equilibrating. Dynamic temperature compensation technology automatically adjusts errors caused by temperature changes to ensure weighing accuracy. The internal calibrate function is triggered by time or temperature and automatically calibrates to maintain accuracy.
Applications
Laboratory, industry, scientific research, education, quality control
Steps
1. Place the Balance on the stable plane and connect the Power Adapter
2. Turn on the power switch and wait for the Balance self-test to complete.
3. Select the weighing unit (e.g. mg, g, etc.) as required.
4. Place the item to be weighed on the pan and wait for the display to be stable
5. Read and record the weighing results
6. Turn off the power after use and clean the pan
Notice
• Balance in a stable, vibration-free environment
• Avoid operating in extreme temperatures (operating temperature 10~ 30 ° C) or humidity conditions
• Make sure the Balance is preheated and complete the self-test before weighing
• Do not exceed Capacity 500g to prevent damage to the Sensor
Periodically calibrated internally to ensure accuracy
• Use a soft cloth when cleaning to avoid liquids from entering the equipment
• Power Adapter input is 110-240V to ensure voltage matching
• Disconnect the power supply when not in use for a long time