Overview
AN AD24 high-speed dispersion Mulser is a laboratory equipment. It is driven by a miniature high-power high-speed motor imported from Germany. It is suitable for Stir mixing of low to medium and high viscosity liquids, dispersion homogenization of granular materials and crushing of tissue cells. The machine is composed of a high-speed motor, a stainless steel dispersion homogenization head and a base. The main engine adopts an engineering plastic and a die-casting aluminum alloy body. The operation is stable and the speed is adjustable to ensure effective and controllable experimental operation.
Features
1. Driven by a miniature high-power high-speed motor imported from Germany.
2. The main engine is made of high-quality engineering plastic and die-cast aluminum alloy, and the operation state is stable.
The working head is made of high-quality stainless steel (SS316) for a wide range of experimental applications.
4. The running speed can be adjusted and controlled (6 gears, 13,000 to 34,000 rpm) to meet the correction requirements for the running speed in the experiment.
It is suitable for liquid materials with a standard processing capacity of 80~ 5000mL and a maximum processing viscosity of 5000cp.
Principle
The device drives the stainless steel working head swirl/spin through a high-speed motor, so that the experimental medium is subjected to the triple action of extrusion, strong Impact and loss of pressure expansion in the material liquid, so as to achieve the refinement and homogenization of the material, and prevent or reduce the delamination with the material liquid.
Applications
Tissue dispersion in biotechnology, sample preparation, enzyme processing in the food industry, industry, cosmetic industry, paints industry, Petrochemicals, etc
Steps
1. Prepare materials according to experimental requirements (processing capacity 80~ 5000mL, viscosity below 5000cp, temperature below 80 ° C).
2. Immerse the working head of the equipment in the liquid to ensure that the immersion height is between 50mm and 165mm.
3. Connect the power supply (AC 220V 50/60Hz).
4. Select the appropriate gear (6 gears in total) through the speed control knob, and start the device for decentralized emulsification.
5. After the experiment is completed, turn off the power, remove and clean the working head.
Notice
Please ensure that the height of the working head immersed in the liquid is between a minimum of 50mm and a maximum of 165mm.
* This machine is suitable for liquid materials with a viscosity below 0.2Pa.s (ie 5000cp) and a temperature below 80 ° C.
* Ambient Temperature should be ≤ 40 ℃, relative humidity should be ≤ 80%.
* Device Duty Cycle is S1 (continuous), do not overload for a long time.
* Power frequency is AC 220V before use.
Do not idle the working head without liquid or insufficient liquid.