Overview
HUITAI DZF-6020 Vacuum oven is widely used in biochemistry, chemical industry, agricultural research, environment protection and other fields for powder drying, baking and disinfection and sterilization of Glass containers. Especially suitable for rapid and effective drying of heat sensitive, easy analysis, easy oxidation substances and complex components.
Features
1. The heating element is located on the Shelf inside the box, which shortens the heating time by 50% compared with the traditional Vacuum oven.
2. Large-screen LCD display, multiple sets of data are displayed on one screen, and the menu-type Operating interface is simple and easy to understand.
3, the use of special technology stamping aluminum shelf, high temperature non-oxidation, heat conduction fast, reduce heat loss.
4. Cuboid studio, with high effective volume utilization, accurate and reliable temperature control by microcomputer temperature controller.
5. Tempered bulletproof double-decked Glass Door to observe indoor objects, can be filled with inert gas.
6, the box door closed elastic adjustable, integrally formed silicone rubber door seal to ensure a high Vacuum Level.
7. The working room is made of stainless steel plate, the box shell is made of Cold rolled steel plate, and the surface is electrostatic spraying.
8, storage, heating, testing and drying in an oxygen-free or inert gas environment, will not cause oxidation.
Principle
The vacuum environment reduces the boiling point of the liquid and makes the heat-sensitive material easy to dry; the vacuum drying method can shorten the drying time, especially for powder or granular samples; the risk of thermal explosion of oxides is eliminated under vacuum or inert conditions, and the powder sample will not be blown by the flowing air.
Applications
Biochemistry, chemical industry, agricultural research, environment protection
Steps
1. Place the sample in the studio.
2. Adjust the closure tightness of the box door as required.
3. Set the microcomputer temperature controller parameters, such as temperature and time.
4. Start the equipment for drying treatment.
5. Remove the sample after drying.
Notice
• Power supply voltage of AC220V 50Hz before use.
• Pay attention to Vacuum Level control during operation to avoid exceeding 133Pa.
• When drying heat-sensitive substances, it should be carried out in a vacuum or inert gas environment.
• Regularly check the door seal and heating elements to ensure the normal operation of the equipment.
• Avoid use in humid or flammable environments.