Overview
BOZHEN BZH-12-12 Plastic glass fiber content determination box type resistance furnace is a new generation of energy-saving intelligent products with high performance mechatronics. It is mainly used to measure plastic glass fiber and Miscellaneous content. It is a satisfactory equipment for plastic processing, production enterprises, scientific research departments, and universities. The equipment can intelligently control time and temperature in sections.
Features
1. Using domestic professional technology, the furnace shell is made of good steel plate by folding welding and drawing, and the control box is integrated with the furnace body
2. Beautiful appearance, avoiding the cumbersome installation of customers. The inner lining is a rectangular integral lining made of high aluminum and silicon carbide refractory materials
3, made of iron-nickel-chromium-aluminum alloy wire wound into a spiral heating element through the inner lining, under, left and right wire groove, up and down left and right heating
4. The furnace is a sealing structure, and the refractory fiber aluminum silicate cotton and diatomaceous earth products are used as insulation layers between the furnace and the furnace shell
5. The overall design is reasonable, the service life is long, and the left opening door and chimney smoke exhaust device are easy to operate. Beautiful appearance, exquisite craftsmanship, sturdy and durable, not easy to deform
Principle
Adopt industrial PID self-tuning control technology, microcomputer intelligent control, after setting the temperature, the instrument controls the heating output by itself, and the temperature control is accurate and stable. With start timing or thermostatic start Timing function after setting the temperature. Using K-indexed nickel-chromium-nickel-silicon Thermocouple for temperature measurement, temperature control is realized through PID adjustment.
Applications
Plastic processing, production enterprises, scientific research departments, universities
Standards
ISO 3451-1:1997, GB/T9345.1-2008 Plastics, Determination of ash, Part 1 General methods
Steps
1. Working time and temperature
2. Boot operation
3. Automatic shutdown after thermostatic time
Notice
• The equipment has an over-temperature alarm device, which automatically cuts off the heating output when it exceeds the temperature
• Power supply voltage 220V 50Hz before use
• Pay attention to the high temperature of the furnace body and be cautious when operating
• Regularly check the status of heating elements and insulation layer