Overview
YIHENG IT-09B5 Magnetic Stirrer is a laboratory equipment with a maximum stirring capacity of 5 liters and a stirring Speed range of 200 to 1800rpm. The product can measure the temperature of the working plate. When the temperature exceeds 55 ° C, the indicator light will remind the user to avoid touching to prevent burns.
Features
1. The LCD screen adopts the menu-type Operating interface, and the parameters such as temperature, Rotation speed and countdown are displayed on one screen, and the operation is simple and easy to understand.
2. Microcomputer control technology is used to ensure that the Rotation speed remains constant when the stirring viscosity changes.
3. The shell is aluminum alloy die-casting, which has the characteristics of high temperature resistance, corrosion resistance and easy cleaning.
4. The countertop is die-cast by aluminum alloy material and sprayed with external enamel, which has good thermal conductivity and corrosion resistance, and the smooth surface is easy to clean.
5. High output torque, suitable for stirring solutions with low viscosity and large capacity or high viscosity and small capacity.
6. The control panel is designed with anti-corrosion liquid conduction tank to prevent liquid overflow and damage to electronic devices.
Principle
IT-09B5 Magnetic Stirrer drives the high temperature resistant strong magnet swirl/spin to generate a swirl/spin magnetic field through a micro-motor, which drives the stirrer in the container to rotate, thereby stirring, mixing or auxiliary heating of the liquid, so that the solution is fully mixed at a set temperature.
Applications
Laboratory, Chemical industry, Biopharmaceutical, Food processing, Education and research
Steps
1. Place the device on a smooth surface and connect it to a power source.
2. Turn on the power switch and set the required temperature, Rotation speed and countdown parameters through the LCD screen.
3. Put the stirrer into the container and place it on the mixing table.
4. Start the mixing function and the equipment starts to work.
5. Monitor the parameters on the display and adjust them if necessary.
6. After use, turn off the power and clean the countertop.
Notice
• When the working plate temperature exceeds 55 ° C, the indicator light flashes to remind, do not touch to avoid burns.
• Avoid direct contact with the mixing table or internal components when the equipment is running.
• Make sure the equipment is powered off when cleaning, and wipe it with a soft cloth.
• Clean up the liquid in time to prevent it from entering the equipment.
• Only for stirring solutions in the specified volume and viscosity range.
• Regularly check the Appearance and function of the equipment, and stop using it if there is any abnormality.