| Product Name | HANNA HI9829-11042 Multi-ParaMeter Detector | LeiCi DZS-706F Multi-ParaMeter Analyzer |
| Product Price | $8494.00 | $1142.00 |
| Details | Details | |
| Model comparison | ||
| Type | Multiparameter | Multiparameter |
| Style | Portable | desktop |
| Detection Principle | Electrode Conductivity DO Method | |
| Measurement parameters | pH DO EC NH₃-N NTU TDS ORP Salinity Temperature | pH DO EC TDS Salinity Temperature |
| Range | Wide Range | Wide Range |
| Accuracy | ±2%FS | ±1%FS |
| Resolution | 1 | 0.1 |
| Data Output | USB | RS232 USB |
| IP Rating | IP68 | IP54 |
| Parameter comparison | ||
| PH Measurement Range | 0.00~14.00pH/±600.0mV | |
| PH resolution | 0.01pH/0.1mV | |
| PH Measurement accuracy | ±0.02pH/±0.5mV | |
| PH calibrated mode | Up to three points of automatic calibrating and one custom calibrating with five standard calibrating points 4.01, 6.86, 7.01, 9.18, 10.01 pH parameters | |
| Measurement Range of oxidation reduction | ±2000mV | |
| Oxidation reduction resolution | 0.1mV | |
| Measurement accuracy of oxidation reduction | ±1.0mV | |
| Oxidation reduction calibrated mode | Single point custom calibrate | |
| Measurement Range of conductivity | 0~ 200mS/cm (Absolute Conductivity detector up~ 400mS/cm) | |
| Conductivity resolution | 1μS/cm、0.001mS/cm、0.01mS/cm、0.1mS/cm、1mS/cm | |
| Measurement accuracy of conductivity | Reading +/- 1% or +/- 1 μS/cm, whichever is the largest | |
| Conductivity calibrated mode | Automatic recognition of single point calibrating, user-defined calibrating, built-in six Conductivity detector standard calibrating point 84 μS/cm, 1413 μS/cm, 5.00mS/cm, 12.88mS/cm, 80.0mS/cm, 111.8mS/cm | |
| Turbidity Measurement Range | 0.0to99.9FNU;100~1000FNU | |
| Turbidity resolution | 0.1σt,σ0,σ15 | |
| Measurement accuracy of turbidity | ±1σt,σ0,σ15 | |
| Turbidity calibrated mode | To calibrate conductivity or salinity | |
| DO Measurement Range | 0.0~500.0%;0.00~50.00ppm | (0.00~20.00)mg/L |
| DOS resolution | 0.1%;0.01ppm | 0.01mg/L |
| DO Measurement accuracy | ≤ 30.00mg/L: reading +/- 1.5% or 0.10mg/L (ppm) whichever is the largest; > 30.00mg/L: reading +/- 3% ≤ 300.0%: reading +/- 1.5% or 1.0%, whichever is the largest; > 300.0%: reading +/- 3% | |
| DO calibrated mode | Single point or double point calibrated Automatic calibration, 100% calibrated automatically in zero or air | |
| temperature compensation | Automatic temperature compensation, -5.00~ 55.00 ° C, 23.00~ 131.00 ° F | |
| GPS function | 12-Channel receiver, within 10m Range (HI9829G series only) | |
| Data storage | 45000 measurement Data storage, recording interval settings: 1 second to 3 hours, USB Interface | |
| Interface | USB Interface (HI9 2000-15 Data Management Package recommended) | |
| Protection Rating | Host: IP67, Dedicated Multi-Parameter Electrode: IP68 | |
| Other functions | FastTrackerTMTAG ID | |
| Power mode | 1.5Vx4AAA battery or 1.2Vx4 rechargeable battery, AC230V/12DC Power Adapter | |
| Operating Environment | 0~50℃(32 to 122℉);RH 100% | Ambient Temperature: (0~ 40) ℃ relative humidity: not more than 85% |
| Weight/Size | Host Size: 221x115x55mm, host weight: 750g | |