| Measurement Range |
0~1000ml/min/0.01μm/(pa.s)~100μm/(pa.s)
|
A file (0~ 1000) ml/min, B file (0~ 5000) ml/min, C file (0~ 10000) ml/min, D file (0~ 100000) ml/min (B, C, D file are optional gears)
|
| test area |
Standard specimen test is 10 +/- 0.2cm ²
|
The standard specimen test is 10 +/- 0.2cm ², and the specimen must be larger than this Size
|
| Differential pressure in test area |
Standard test pressure difference is 1 +/- 0.01kpa
|
The standard test pressure difference is 1 +/- 0.01kpa, and special materials can be customized for other pressure differences as required
|
| Measurement accuracy |
< 100Ml, error 1ml; > 100ml, error 5ml
|
< 100Ml, error 1ml; > 100ml, error 5ml
|
| Inner diameter of clamp ring |
Standard specimen clamping inner diameter 35.68 +/- 0.05mm
|
Standard specimen clamping inner diameter 35.68 +/- 0.05mm
|
| Coaxiality of upper and lower clamp ring center hole |
Strictly control coaxiality deviation < 0.05mm
|
Strictly control coaxiality, deviation < 0.05mm
|
| Environmental conditions temperature |
(0~ 35) ℃, Humidity < 85%
|
(0~ 35) ℃, Humidity < 85%
|