| Volume in the box |
≥ 0.5 cubic meters with Glass observation door
|
|
| Glowing wire temperature |
Adjustable at room temperature~ 1000 ° C, temperature measurement adopts Φ 1.0mm nickel-chromium armor (K Type) imported Thermocouple
|
500~ 1000℃, adjustable digital display
|
| Temperature tolerance error |
500 ~ 750℃±5℃;750 ~ 960℃±10℃
|
+/- 10℃ (500~ 750℃), +/- 1 5 ℃ (750~ 1000℃), the temperature is the meter degree +/- 0.5 level
|
| The force exerted by the hot wire on the sample |
1.0 +/- 0.2N, with a press-in depth of 7mm +/- 0.5mm
|
|
| Glowing wire |
The glowing wire is a ring of standard Size made of nickel/chromium (80/20) wire with a diameter of 4mm; the diameter of the glowing wire is 4.0 +/-0.04mm before bending, and it must not be reduced to 90% after bending. Miscellaneous dimensions are made according to the standard error requirements
|
4.0Mm diameter rings made of nickel/chrome wire
|
| test data |
Touch screen display and print
|
|
| Timer |
Adjustable in the Range of 0~ 99 minutes and 99 seconds, accuracy +/- 1 second, burning time (Ta) can be set, ignition duration (Ti) and flame extinguishment time (Te) can be recorded
|
|