| Volume in the box |
≥ 0.5 cubic meters with Glass observation door
|
|
| electrical load |
The test Voltage is adjustable between 100 and 600V. When the short-circuit current is 1A +/- 0.1A, the Voltage drop should not exceed 10%. When the short-circuit leakage current in the test circuit is equal to or greater than 0.5A, the time is maintained for 2 seconds. The relay operates to cut off the current, indicating that the test product is unqualified
|
|
| Bipolar force on the sample |
Electrode with rectangular platinum plating, the force of the two poles on the sample is 1.0N +/- 0.05N
|
|
| Dripping device |
Drop height adjustable from 30mm~ 40mm, drop Size 44~ 50 drops/cm ³, drop time interval 30 +/- 1 second
|
The dripping time interval can be set arbitrarily
|
| Control mode |
PLC + Touchscreen control for manual/automatic control, parameter setting and display, and operation monitoring, etc. Easy to operate and easy to read real-time experimental data, and print experimental results (optional)
|
|