Product Name HUDA QD-E708-1000K40-X15 Rapid Rate Temperature Chamber kesionots KS-HD1000LC low/high temperature test chamber
Product Price $28119.00 In coming
Details Details
Model comparison
Test Type High/Low Temp Temp Cycling High/Low Temp
Min-TEMP -40℃ -40℃
Max-TEMP 150℃ 150℃
working volume 801-1000L 801-1000L
Temp Fluctuation ±0.5℃ ±0.5℃
Temp Uniformity ≤2.0℃ ≤1.0℃
Temp Deviation ±1.0℃
Temp Change Rate 3-5℃/min
Cooling Method Water-cooled Water-cooled
Parameter comparison
Temperature Range -40~+150℃ -40℃~150℃
Moisture control Range 20~98%R.H
Interior Volume 1000L 1000L
Effective Size of inner box W1000*D1000*H1000mm
Dimension About W1550 * H2330 * D2480mm 150*186*157cm
weight About 1200 kg
Maximum power 60Kw
Maximum current 85A
Power supply conditions and power supply AC 380V three-phase four-wire system + protective grounding N; Voltage allowable fluctuation Range 10% V; frequency allowable fluctuation Range 50 +/- 0.5HZ; TN-S mode power supply or TT mode power supply; protective grounding resistance less than 4Ω
Temperature resolution 0.01℃ ±0.1℃
Humidity resolution 0.1%R.H
Temperature deviation ±1℃
Humidity deviation ±2%R.H
Temperature Uniformity ≤2℃ ±1℃
Humidity uniformity ±3%R.H
Temperature Fluctuation ±0.5℃ ±0.5℃
humidity fluctuation ±2%R.H
linear cooling rate 15 ℃/min (under no-load conditions)
Refrigeration compressor Low temperature compressor imported from Germany
Dehumidification compressor Full enclosed compressor dehumidification
Chill down Blisters chill down
evaporator finned tube heat exchanger
Evaporative condenser A plate heat exchanger brazed with high temperature copper has a small surface area, small energy loss, and high heat exchange efficiency
Throttle device Thermal expansion valve, capillary tubing
Control mode of refrigerator The programmable logic controller of the control system automatically adjusts the operating conditions of the refrigerator according to the test conditions, the compressor return air chill down circuit, and the energy pressure regulating valve
refrigerant Using an environmentally friendly refrigerant R404a/R23 with an ozone depletion index of 0
Chill down the water tower Low noise air cooled down water tower (optional)
circulating water pump Low noise pipeline pump powerful water supply circulation (optional)
controller Imported controller to control temperature and humidity separately Korean "TEMI" or Japanese "OYO" card to choose from
RS-232 communication interface Can be connected to a computer to display Linear dispersion, data acquisition; can be used as a monitoring and remote control system; can be synchronized with multiple machines
SD memory card For storing test data and Linear dispersion
setting method Chinese/English interface, touch input
operation mode Program method/valuation method
setting Range Temperature Range 5 ℃
Display resolution Temperature: 0.01 ℃; Humidity: 0.1% R.H; Time: 1min
Power-off memory function The power failure recovery mode can be set as: hot start/cold start/stop
Boot reservation function You can set the boot time at will, and the machine will run automatically after turning on the power supply.
Temperature measurement body PT100 Platinum Resistor
Heating system Using nickel alloy heating tube heating, circulating fan strong air circulation, P.I.D control heating to achieve temperature equilibrating
Humidification and dehumidification system Using electric boiler humidification mode, the compressor starts dehumidification, and P.I.D controls the humidification amount to reach the required humidity
Linear dispersion recording function With battery-protected RAM, it can save the setting value, sampling value and sampling time of the device; the maximum recording time is 60 days (when the sampling period is 1.5min)
Software use environment IBM PC compatible machine, CPU above PII, memory above 128M, Simplified Chinese Windows2000 or Simplified Chinese WindowsXP operating system
input platinum resistance
subsidiary function Fault alarm and cause, processing prompt function; power failure protection function; upper and lower limit temperature protection function
Water quality requirements It is required to meet the second-class water quality of the National Analytical Laboratory, and the solid impurity content of the water is less than 1mg/L
Water supply method Automatic/manual dual water supply mode
water supply device 1. The machine is equipped with a small water softener filtrate unit, which only needs to be connected to tap water, and the pure water required for automatic production spray is required to effectively prevent impurities in the water quality from adhering to the surface of the sample and affecting the test results; water supply pressure: 0.1~ 0.4Mpa; 2. PVC large water tank, which can be manually replenished when water supply is inconvenient
Refrigeration system protection device Compressor overheating, overcurrent, overpressure, phase inversion, oil shortage protection
Test chamber protection device Ultimate overtemperature, fan, motor overheating relay
Encasement Encasement boxes meeting the requirements of QB/BWD008-2001
mode of transportation Whole machine freight
acceptance criteria The technical indicators shall be agreed upon by both parties
acceptance method Formal acceptance at the buyer's site
Use environmental conditions Environment temperature is 5~ 30℃, relative humidity ≤ 85% R.H
Storage environment The environment temperature of the equipment should be kept within 0 ° C~ + 45 ° C; when the environment temperature is lower than 0 ° C, (the equipment stops for a long time) the water left in the equipment should be discharged clean to avoid the water in the pipeline freezing and damaging the pipeline