| Product Name | Apkjgroup 2AP-CJ-80B Thermal Shock Test Chamber | AISRY NBC-80B Thermal Shock Test Chamber |
| Product Price | $17985.00 | $15403.00 |
| Details | Details | |
| Model comparison | ||
| Structure Type | Two-zone | Two-zone |
| Sample Movement | Movable (Basket) | Movable (Basket) |
| Shock Temp Range | -55℃~+150℃ | -40℃~+150℃ |
| Hot Zone Range | +60℃~+180℃ | |
| Cold Zone Range | -10℃~-65℃ | -10℃~-55℃ |
| volume | 51-100L | 51-100L |
| Recovery Time | ≤5min | |
| Transfer Time | ≤10s | ≤10s |
| Cooling Method | Air-cooled | Air-cooled |
| Sample Load Capacity | 21-50kg | 11-20kg |
| Defrost Function | Automatic Defrost | |
| Parameter comparison | ||
| Temperature Range | -55 ℃~ + 150℃ (air-cooled) | |
| Temperature Fluctuation | ±0.5℃ | ±2℃ |
| Temperature deviation | ±2.0℃ | |
| weight | About 550kg | |
| content product | 80L | |
| Internal dimensions | 400*500*400mm (W*H*D) | |
| External dimensions | 1400*1950*1850mm (W*H*D) | |
| Low Temperature Impact Range | -10℃~ -55 ℃ (optional) | |
| High Temperature Impact Range | 60℃~ 150℃ (arbitrarily adjustable) | |
| Test environmental conditions | The surrounding environment temperature of the machine is maintained between + 25~ + 30 ° C, relative humidity: ≤ 85%; air pressure: 86kPa~ 106kPa normal atmospheric pressure | |
| High Preheating Temperature Range | RT~+195℃ | |
| heating time | + 50℃ → + 195 ℃ ≤ 40min; Note: The heating time is the performance of the high greenhouse when running alone | |
| Temperature Range of low greenhouse precooling | RT~-65℃ | |
| cooling time | + 20℃ → -65 ℃ ≤ about 90min; Note: The cooling time is the performance of the low greenhouse when running alone | |
| Test method | Place the product in the hanging basket test area, two boxes of mobile Thermal Shock, safe air pressure to drive the test object up and down Impact | |
| temperature recovery time | ≤ 5min; the conversion temperature only needs ≤ 10s (the movement time of the hanging basket: within 3 seconds) | |
| Impact exposure time | ≥30min | |
| Outer wall material | The outer material can be selected from SUS304 high-quality stainless steel or fine powder paint treatment | |
| Inner wall material | The inner plate material is high-quality SUS304 stainless steel, with good corrosion resistance, heat resistance, low temperature strength and mechanical characteristics; stamping, bending and other hot workability is good, no heat treatment hardening phenomenon (no magnetic, use temperature -196 ℃~ 800℃) | |
| Box insulation material | Rigid polyurethane foam | |
| Air conditioning channel | Fan, heater, evaporator, damper, temperature sensor | |
| Specimen shelf | 1 set, a Total of two layers (with four hooks) bearing 25kg (please indicate when placing an order for larger bearing) | |
| Lead Hole (Optional) | 1, located on the left side of the box, with a round hole with a Size of Φ 50mm | |
| Adjusting feet and adjusting wheels | Adjusting feet: 4, for supporting the test chamber; Adjusting wheels: 4, for moving the test chamber | |
| Test chamber door | Single opening, left hinge, right handle | |
| circulating fan | High temperature/ambient temperature exposure: 3Φ, 400W, centrifugal; low temperature exposure: 3Φ, 400W, centrifugal | |
| Control Panel | Controller display, fault indicator, timer, USB or R232 interface, over-temperature protection setter, main power switch | |
| Machinery room | Refrigeration unit, connected drainage device, exhaust fan | |
| Distribution control cabinet | Total power leakage circuit breaker, controller, distribution board, cooling fan, high and low temperature circulating fan motor | |
| heater | Nickel chromium alloy heating wire heater | |
| Control mode of heater | No touchpoint equal period pulse width modulation, SSR (solid state relay) | |
| Power cord holes and drainage holes | Located on the back of the box | |
| Refrigeration compressor | Original imported France Taikang or Japan Hitachi fully enclosed low noise rotary compressor | Adopt imported compressor unit |
| evaporator | finned tube heat exchanger | |
| condenser | shell and tube heat exchanger | |
| Throttle device | Thermal expansion valve, capillary tubing | |
| Evaporative condenser | Stainless steel brazed plate heat exchanger | |
| Control mode of refrigerator | The PLC (programmable logic controller) of the control system automatically adjusts the operating conditions of the refrigerator, the compressor return air chill down circuit, the energy adjustment circuit, and the condensing pressure control valve according to the test conditions | |
| refrigerant | R404a/R23 (ozone depletion index is both 0) | |
| controller | Programmable controller, touch screen input, can arbitrarily set different high and low temperature Impact temperature points according to customer requirements, to meet the different needs of testing | |
| display | LCD liquid crystal display | |
| operation mode | procedural approach | Program running, constant running |
| setting method | All Chinese LCD display (Chinese and English can be switched) LCD touch panel (resistant to 100.000 line drawing test) | |
| program capacity | Up to 120 groups, each program is divided into three zones, three zones and three segments. Each command can be up to 999 times; the program segment time is set to 0~ 999Hour59min | |
| Display resolution | Temperature: 0.1 ℃; Time: 0.1 min | |
| electrical input | Thermocouple | |
| Storage device | USB and R232 communication interface | |
| Control mode | Anti-integral saturation PID, fuzzy algorithm; BTC equilibrating temperature control mode | |
| Linear dispersion recording function | Real-time display of graphics during program execution Linear dispersion | |
| subsidiary function | With power-off program memory, automatically start and continue to execute program functions after power-back, with date, time adjustment, reservation start, shutdown function and screen lock (LOCK) function | |
| environmental conditions | Temperature: 5 ℃~ 35 ℃; relative humidity: ≤ 85%; air pressure: 86kPa~ 106kPa normal atmospheric pressure | |
| Air source pressure | (5~7)kg/cm² | |
| Air source connecting pipe diameter | Φ8mm | |
| Air source flow | 0.1m³/min | |
| Maximum current consumption | Current: 18A; Power: about 12.0Kw [Voltage Range: no more than +/- 10%] | |
| Requirements for storage environment | When the equipment is not working, the temperature of the environment should be kept within 0 ° C~ + 45 ° C; when the environment temperature is lower than 0 ° C, the water remaining in the equipment should be drained clean to avoid freezing water in the pipeline | |