Product Name HANNA HI2003-02 Benchtop Conductivity Meter HANNA HI2004-02 Benchtop dissolved oxygen Meter
Product Price $1264.00 $1300.00
Details Details
Model comparison
Type Desktop
Range Miscellaneous
Precision Miscellaneous
Resolution Miscellaneous
Temperature compensation Automatic
PH Without
TDS With
Salinity with
Resistivity Without
Accessories
Parameter comparison
Resolution-EC 0.01μS/cm、0.1μS/cm、1μS/cm、0.01 mS/cm、0.1 mS/cm
Accuracy @25 ℃ -EC +/- 1 % +/- ( 0.05 μS/cm or 1 word), whichever is larger
Calibrated mode -EC Electrode drying state, 0.00μS/cm Automatic calibration in the air; built-in standard calibrating point single point automatic identification calibrating; built-in calibrating point: 84μS/cm, 1413μS/cm, 5.00mS/cm, 12.88mS/cm, 80.0mS/cm, 118.8mS/cm
Measurement Range - TDS 0.00 to 14.99 mg/L (ppm), 15.0 to 149.9mg/L (ppm), 150 to 1499mg/L (ppm); 1.50 to 14.99g/L, 15.0 to 100.0g/L; up to 400.0g/L (actual/absolute Total solids solubility, closing temperature compensation), TDS conversion factor: 0.80
Resolution-TDS 0.01mg/L(ppm)、0.1mg/L(ppm)、1mg/L(ppm)、0.01g/L、0.1 g/L(ppt)
Accuracy @25 ℃ -TDS +/- 1 % +/- ( 0.03mg/L or 1 word), whichever is greater
Calibrated Mode - TDS Conductivity detector calibrated without Total solids solubility
TDS conversion factor Adjustable from 0.45 to 0.80 (default 0.5)
Measurement Range of salinity Actual salinity: 0.00~ 42.00 PSU, seawater salinity: 0.00~ 80.00ppt (g/L); salinity: 0.0~ 400.0% NaCl
Salinity resolution 0.01 PSU、0.1g/L、0.1%
Salinity accuracy @25 ℃ Reading +/- 1%
Salinity calibrated mode PSU and g/L salinity units do not need to be calibrated according to conductivity calibrating results;% units are optional HI7037 salinity standard buffer (100%) for single point calibrating
Measurement Range -20.0~120.0℃、-4.0~248.0℉
temperature resolution 0.1℃、0.1°℉
Measurement accuracy ±0.5℃、±0.9°℉ ±0.5℃; ±0.9℉
Electrode Type HI763100 Customized built-in temperature sensor EC/TDS dedicated Electrode, Interface Type: 3.5mm, 1 meter wire length
temperature compensation Automatic temperature (ATC) compensation, -5~ 100 ° C (32~ 212 ° F); when measuring absolute conductivity, select off temperature compensation (NOTC)
compensation coefficient Conductivity temperature coefficient Range: 0.00~ 6.00%/℃
GLP Calibrated data includes information such as date, time and calibrated point
storage mode Automatically store 1000 sets of measurement data, store 200 sets of measurement data at intervals; manually store 600 sets of measurement data
Interface Dual communication interface mode: USB Interface and micro Interface
Power mode Built-in rechargeable battery, 5VDC Power Adapter and special charging multi-function bracket
Operating Environment 0 to 50℃(32 to 122 F)
Size weight Host Size: 202x140x12.7mm, host weight: 250g