Overview
The Netherlands TQC VF2087 No. 4 Ford viscosity cup is a portable immersion Ford viscosity cup, designed according to ASTM D1200 standard, suitable for production site viscosity testing. The product has an outlet diameter of 4.1mm, a solution viscosity measurement range of 70~ 370cst, and a flow time of 30~ 100s.
Features
1. Titanium anodic oxidation aluminum cup material, stainless steel filter and handle material to ensure durability
2. Outflow port diameter 4.1mm, precise control of fluid flow
3. Viscosity measurement range 70~ 370cst, flow time 30~ 100s, covering a wide range of application requirements
4. Portable immersion design for easy on-site operation
Principle
The VF2087 Ford viscosity cup determines the viscosity of a liquid by measuring the time it takes for a specific volume of liquid to flow out of the standard outlet. According to the ASTM D1200 standard, at 25 +/- 0.2 ° C Ambient Temperature, the time for the liquid to completely flow out of the cup is recorded, which is proportional to the liquid viscosity.
Applications
Production site viscosity testing, fluid property measurement
Steps
1. Dip the Ford viscosity cup into the liquid to be tested
2. Ensure Ambient Temperature is maintained at 25 +/- 0.2 ℃
3. Lift Ford viscosity cup and start timing
4. Record the time for the liquid to completely flow out of the outlet
5. Calculate the liquid viscosity value based on the outflow time
Notice
• Ensure Ambient Temperature Stable at 25 +/- 0.2 ℃ Before Use
• Keep the outlet clean to avoid clogging
• Avoid vibration interference during measurement process
• Clean and maintain in time after use