$ 864.00Ex works
Equipped with 7-inch TFT display, supports 5 ion Measuring methods and a variety of concentration units, automatically recognizes ion selective ElectRode, can store 1000 sets of data, Measurement range 5 × 10~ 1M.
Category : Ion Meter
Brand : Bante
SKU : NB014766
Service :The delivery and service By NBCLab.Com
The A131-Ca Precision Lab ion meter is a professional instrument of the BANTE brand, equipped with a 7-inch TFT display, supports 5 ion Measuring methods, and the system menu contains 13 optional parameters. The meter is suitable for the measurement of Ion concentration and water quality Hardness of liquids, providing High Accuracy measurement and a variety of practical functions.
1.7-Inch TFT display, clear and intuitive display. 2. Optional 5 ion Measuring methods: direct reading method, known addition method, known reduction method, sample addition method, sample reduction method. 3. Choose from a variety of concentration units: ppm, mg/L, mol/L, mmol/L. 4.2 to 5 point button calibrated, supports 8 concentration calibrated points. 5. Automatically recognizes ion-selective Electrodes without specifying the ion type. 6. Electrode management function, can store 3 ion Electrode slope and call calibrated data. 7. The calibrated report provides standard and Electrode slope details. 8. Water Hardness mode supports 2 to 5 point button calibrating, optional units include German degree, British degree, French degree, mmol/L, mg/L. 9. calibrate expiration reminder function. 10. Automatic end point locking, retaining stable Measured value. 11. Interval reading mode, regularly send data to PC or printer. 12. Password protection to prevent unauthorized operation. 13. Extended memory to store up to 1000 sets of data. 14. USB Communication interface, easy to connect peripherals. 15. One-click reset function to restore factory default settings.
The ion meter detects Ion concentration in liquids by ion-selective Electrode. Measuring methods such as direct reading and addition are supported to automatically identify the Electrode type and calculate the slope. For measurement, the meter processes the Electrode signal, combined with temperature compensation (manual or automatic), to display Ion concentration or Water Hardness values, and can stabilize the reading with data locking.
Laboratory Ion concentration measurement, Water Hardness analysis, environment monitoring, industrial process control, scientific research and education
1. Connect the Power Adapter and turn it on. 2. Select Ion or Water Hardness Measurement mode. 3. Perform 2 to 5 point key calibration, from low to high concentration. 4. Insert Ion Selective Electrode or Temperature Probe. 5. Measurement parameters such as units and stable conditions are set. 6. Start measurement, the meter automatically locks the end point. 7. View and store data, which can be exported via the USB interface. 8. Use the one-click reset function to restore the default settings (if necessary).
• Regularly calibrate and set the calibrate reminder function. • Check the Electrode status before use to ensure that the automatic identification is normal. • Avoid unauthorized operation and enable password protection. • Make sure the temperature compensation is set correctly during measurement. • Memory capacity when storing data and export it in time. • Use specified Accessories such as ionic standard solution and temperature probe. • Keep the meter clean and avoid violent vibrations.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.