Overview
The YUE FENG FP6410 Flame photometer is a compact, delicate and beautiful analytical instrument that uses liquefied petroleum gas as fuel, has a dual-channel LED digital display function, can detect potassium and sodium at the same time, and has a lithium calcium detection function.
Features
1. Appearance Small, exquisite and beautiful
2. Dual-channel LED digital display, 4-digit digital display per channel, clear reading
3. Unique ignition gas path, Keypad Operation automatic ignition, fast and reliable ignition
4. Potassium and sodium can be detected simultaneously, with the detection function of lithium calcium
5. Silicon Photocell, photovoltaic cell detection, accurate and stable measurement results
6. Small Sample Size, High Sensitivity
7. Using liquefied petroleum gas as fuel, easy to use
8. With flameout protection function, safe and reliable performance
9. The flame can be adjusted to a good state at one time, making it easy to operate
Principle
The Flame photometer uses Interference filter as Measurement principle, through the Silicon Photocell, photovoltaic cell detection sample in the flame excitation generated characteristic Light spectrum, to achieve quantitative analysis of potassium, sodium and other elements.
Applications
Chemical analysis, laboratory testing, environment monitoring, medical inspection, industrial quality control
Standards
Stability standard: the relative maximum change of instrument indication within 15S ≤ 3%, Linearity standard: potassium K ≤ 0.005mmol/L, sodium Na ≤ 0.03mmol/L, detection limit standard: potassium K ≤ 0.004mmol/L, sodium Na ≤ 0.008mmol/L, Response Time standard: < 8S
Steps
1. Prepare samples and LPG fuel
2. Turn on the instrument power
3. Press the automatic ignition button to ignite.
4. Adjust the flame to the optimum state
5. Introduce the sample into the flame for testing
6. Read potassium and sodium concentration data through dual-channel LED display
7. Turn off the instrument after the test is completed
Notice
• Please ensure the normal supply of liquefied petroleum gas before use
• Pay attention to safety when igniting to avoid flame leakage
• Instrument Ambient Temperature should be kept within the range of 10~ 35 ℃
• Regularly check whether the flameout protection function is normal
• The sample volume should be controlled within the appropriate range to ensure Measurement accuracy
• The volume of the instrument is 650 * 400 * 375mm, please make sure there is enough installation space