Please contact customer service
The highest temperature can reach 1200 ℃, temperature control accuracy +/- 1 ℃, the Furnace adopts aluminum oxide microcrystalline fiber material, which has the characteristics of Balanced temperature field, low surface temperature and fast temperature rise and fall rate.
Category : Muffle Furnace
Brand : Shengli
SKU : NB018056
Service :The delivery and service By NBCLab.Com
The SLX1200-50 high temperature furnace is a medium-sized box-type high temperature furnace with a maximum temperature of 1200 ° C. It adopts double-decked shell structure, the furnace uses aluminum oxide microcrystalline fiber material, and is equipped with a 30-stage programmed temperature control system, with a temperature control accuracy of +/- 1 ° C. The furnace has the advantages of Balanced temperature field, low surface temperature, fast temperature rise and fall, and energy saving. It is suitable for high temperature sintering, metal annealing, mass testing and other operations in universities, scientific research institutes and industrial and mining enterprises.
1. The double-decked shell structure is adopted, and the furnace shell is equipped with an air-cooled circulation system to effectively reduce the surface temperature. 2. The furnace adopts aluminum oxide microcrystalline fiber material, which has good thermal insulation performance. 3. Equipped with 30-stage program temperature control system, high temperature control accuracy (+/- 1 ℃). 4. The heating rate is fast (≤ 20 ° C/min), and the temperature field is Balanced. 5. Operating temperature 1200 ℃, can work continuously at 1100 ℃. 6. The furnace volume is large (45L), and the heating zone length is 440mm.
The furnace uses a resistance wire as the heating element. The temperature in the furnace is monitored in real time through a K-type Thermocouple, and the signal is fed back to a 30-stage programmed temperature control system. The temperature control system precisely controls the power output of the heating element to achieve the set heating, holding and cooling procedures. The air-cooled circulation system between the double-decked furnace shells helps to dissipate heat and ensure that the furnace surface temperature is maintained within a safe range.
High temperature sintering, metal annealing, mass testing in universities, research institutes, industrial and mining enterprises
The information does not mention
1. Check whether the Power supply voltage is 380V 50Hz and ensure that the equipment is well grounded. 2. Put the sample to be processed into the furnace (size: 500 * 300 * 300mm). 3. Close the furnace door and make sure the sealing is good. 4. Set the required temperature program (including heating rate, target temperature, holding time, etc.) on the 30-stage programmed temperature control system. 5. Start the heating program and the device starts to run according to the set program. 6. After the program runs, the furnace temperature can be naturally cooled down to a safe temperature before opening the furnace door to take out the sample.
Equipment must be operated by professionals and operating procedures must be strictly followed. * Power supply voltage 380V 50Hz before power on, and check whether the line connection is firm. When the furnace body is working and just after the work, the surface of the furnace shell may be high temperature, please do not touch it to prevent burns. It is strictly forbidden to heat flammable, explosive and toxic items. * When the sample is placed in the furnace, it should be ensured that there is a sufficient distance from the heating element to avoid short circuit or contamination of the furnace. During the operation of the program, do not open the furnace door at will to avoid affecting the temperature field and causing heat loss. When the equipment is not used for a long time, the power supply should be cut off and the inside of the furnace should be kept clean and drying. Check components such as Thermocouple regularly to ensure they are working properly.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.