Overview
HUITAI 4-10T ceramic fiber muffle furnace is a heating auxiliary equipment for dry pretreatment of samples in analytical laboratories, melting experiments in metallurgical laboratories, annealing and quenching experiments in heat treatment departments, and Miscellaneous high temperature applications. The product adopts an integrated design, which optimizes the integration of the furnace body and the control part, and significantly reduces the occupied space.
Features
1. Integrated design of the whole machine, large-screen LCD display, multiple sets of data displayed on one screen, beautiful and elegant, simple operation.
2. Intelligent PID High Accuracy Temperature Control for Power Loss Free.
3. Imported HRE ultra-high temperature alloy heating element, long service life, effective energy saving.
4. The heating rate is fast, and the room temperature rises to 1200 ° C for less than 30 minutes.
5. Less heat pollution, the use of new ceramic fiber insulation materials, furnace body and shell air insulation structure, low surface temperature, heating up to 1000 ℃ to maintain 1 hour after the shell about 50 ℃.
6. Temperature control precision, a variety of thermostats optional, keep the state temperature fluctuation is small (temperature control accuracy +/- 1 ℃, Temperature Uniformity +/- 5 ℃), P-type multi-step thermostat can preset each step temperature rise and fall slope.
Principle
The muffle furnace generates high temperature through resistance wire heating elements, and uses intelligent PID controllers to precisely adjust Power for efficient temperature control. The furnace uses ceramic fiber material, combined with air insulation structure, to reduce heat loss and ensure Temperature uniformity and Stability.
Applications
Analytical laboratories, metallurgical laboratories, heat treatment departments, and high-temperature experimental occasions
Steps
1. Connect the power supply to ensure that the equipment is safely grounded.
2. Turn on the power switch, and the large-screen LCD display starts.
3. Set the required temperature parameters and use the P-type multi-step thermostat to preset the temperature rise and fall slope.
4. Start the heating program and monitor the temperature to rise to the target value.
5. After the experiment is completed, turn off the heating, and turn off the power after the furnace body is chilled down.
Notice
• Check the power supply and grounding before operation to ensure safety.
• Avoid touching the furnace body during the heating process to prevent burns.
• Set the temperature not to exceed the maximum temperature of 1000 ° C.
• Keep the furnace clean to avoid foreign objects affecting performance.
Equipment chilled down before cleaning or maintenance.