Overview
RONGHUA INSTRUMENT SX2-10-12 box type resistance furnace has a maximum Operating temperature of 1200 ℃ and a Rated power of 10KW. It is suitable for elemental analysis and determination in laboratories, industrial and mining enterprises, R & D institutions, and heating treatment such as quenching, annealing, and tempering of small steel parts. The resistance furnace is equipped with a temperature controller and Thermocouple, which can realize temperature indication, regulation and control.
Features
1. Operating temperature up to 1200 ℃.
2. Provide a temperature controller and a temperature measuring thermocouple to achieve temperature indication and adjustment.
3. Heating time is fast, empty furnace to 1200 ℃ less than 80 minutes.
4. The furnace size is 400 * 250 * 160mm, suitable for small workpiece processing.
Principle
The box type resistance furnace converts electric energy into heat energy through the internal electric heating element to heat the furnace. The matching temperature controller receives the temperature signal measured by the Thermocouple in the furnace, and controls the working state of the electric heating element by adjusting the power output, so as to realize the accurate indication, adjustment and stable control of the temperature in the furnace.
Applications
Elemental analysis and determination in laboratories, industrial and mining enterprises, R & D institutions, quenching of small steel parts, annealing of small steel parts, tempering of small steel parts
Standards
Not mentioned in the data
Steps
1. Put the workpiece into the furnace.
2. Set the target temperature and working procedures through the matching temperature controller.
3. Start the resistance furnace to start heating.
4. After the heating process, the workpiece is treated according to the regulations or the furnace is chilled down.
Notice
• Rated voltage of the equipment is 380V three-phase electricity, and it is necessary to ensure that the power supply matches.
• Operating temperature is 1200 ℃, do not use over temperature.
• Power loss of the empty furnace is not more than 2.4KW, please pay attention to the operating energy consumption.
• Follow heat treatment and equipment safety procedures during operation.