$ 2303.00Ex works
With circulating pump monitoring and peristaltic pump dosing system, pH Measurement accuracy +/- 0.05pH, ORP accuracy +/- 5mV, support automatic temperature compensation and data recording functions.
Category : Water Quality Tester
Brand : HANNA
SKU : NB020698
Service :The delivery and service By NBCLab.Com
The BL121-10 is a pool controller under the HANNA brand that is part of the BL120 and BL121 family of products and is designed for the maintenance of sanitized water quality in swimming pools, hot tubs and spa facilities. The system is available in two configurations: the basic version is an in-line model that directly mounts probes and chemical injection fittings to existing piping; and the panel mounted version is equipped with a bypass flow battery that allows the Probe to be calibrated and maintained without shutting down the recirculation pump.
1. Used to maintain the water quality of swimming pools, hot baths, and hot springs 2. Circulation pump monitoring function 3. Display Multi-Parameter simultaneously 4. Customizable alarm reminder system 5. Peristaltic pump dosing system
The controller monitors water quality parameters in real time through pH, ORP and temperature sensors, using automatic temperature compensation and calibrating technology to ensure measurement accuracy. The system automatically adds chemicals through a peristaltic pump according to the set point, and intelligently controls with alarm filtrate options and relay outputs. Data is recorded and transmitted in .csv format via USB flash drives.
Swimming pool, hot tub, spa facilities
GLP standard
1. Install probe and chemical injection fittings into the pipe or use bypass flow batteries 2. Perform pH and ORP calibrating: pH supports single-point Manual calibration or two-point automatic identification calibrating, ORP is single-point Manual calibration 3. Set dosing parameters: adjust pH, ORP setting point and ratio band, configure delayed start and overdose protection 4. Configure the alarm system: enable/disable parameter alarms, set alarm relay controls 5. Select Pump Control Mode: Automatic or Manual, Adjust Flow rate from 0.5 to 3.5 L/h 6. Data Management: Transfer data via USB flash drive, Automatically record Measured values and events 7. Set up password protection: protect settings, calibrate and log call functions
• The power requirements are AC100-240V, Power 10VA, to ensure stable power supply • Ambient Temperature 0~ 50 ° C, maximum relative humidity 95% and no condensation The alarm system is triggered based on 5 consecutive readings, and the alarm value needs to be set reasonably • Use built-in standard buffers (4.01, 7.01, 10.01 pH) when calibrating pH Auxiliary inputs are used to monitor low liquid levels in acid/alkali tanks and chlorine tanks, as well as to maintain input • Data storage period is 60 days, recording interval is 10 seconds, and data is regularly backed up • Probe calibrated during installation and maintenance to avoid affecting water quality monitoring accuracy
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.