Ceramic fiber Furnace, good heat insulation and light weight; PID temperature control accuracy of +/- 1 ℃, Heating time ≤ 30 minutes; Furnace door open automatic power off, with over temperature limit and other safety functions.
Category : Muffle Furnace
Brand : LICHEN
SKU : NB022461
Service :The delivery and service By NBCLab.Com
LICHEN SX2-4-10LT ceramic fiber box type resistance furnace shell is welded with high-quality Cold rolled steel plate, and the furnace door is a side open structure. The interior is insulated by aluminum silicate fiber cotton, which has large heat storage capacity, fast heating and low energy consumption. The product is suitable for elemental analysis and determination in laboratories, industrial and mining enterprises, R & D institutions, etc., and metal heat treatment heating such as small steel parts.
1. Adopting PID temperature control technology, the LCD screen clearly displays the set temperature and current temperature. 2, the use of ceramic fiber furnace, heat insulation effect is good, light weight; furnace control conjoined design, easy to operate. 3, the wire inlaid in the inner wall of the furnace, furnace temperature uniform, heating fast, time-saving; furnace door sealing good, less heat loss, energy saving. 4. The furnace door is made of stainless steel, which is resistant to corrosion and aging; when the furnace door is opened, the heating power supply is automatically disconnected to ensure safe operation. 5, cold-rolled steel shell by electrostatic spraying and fine matte treatment, more corrosion resistance, aging resistance. 6. Standard over-temperature self-limiting, self-setting, door-opening and heat-breaking functions; SX2-LTP models come standard with 30-stage program temperature control.
The resistance furnace is heated by electrical resistance wire as a heat source. Heat accumulates and conducts in the furnace with aluminum silicate fiber cotton insulation to form a uniform high temperature environment. The PID temperature control system monitors the temperature in the furnace through the Sensor and adjusts the Heating power to achieve accurate temperature control.
Elemental analysis and determination in laboratories, industrial and mining enterprises, R & D institutions, small steel parts and Miscellaneous metal heat treatment heating
1. Check the power supply and connection of the equipment to ensure it is in a safe state. 2. Put the sample to be processed into the furnace working area. 3. Close the furnace door and set the required temperature and program through the LCD screen (if it is a programmed temperature control model). 4, start heating, the equipment starts to heat up. 5. When the heating process is over or needs to be interrupted, the furnace door can be opened (the heating power will be automatically disconnected) to take out the sample. 6. After use, turn off the power of the device.
• When the furnace door is opened, the heating power will be automatically disconnected, but the furnace may still be in a high temperature state. Wear protective equipment during operation to prevent burns. • Do not operate beyond the recommended operating temperature (500~ 950 ° C) to avoid damage to the equipment. • When placing the sample, make sure it is within the working size range (300 × 200 × 120mm) and does not touch the heating element. Equipment should be placed on a well-ventilated and stable surface, away from flammable and explosive items. • Regularly check the power cord and equipment Appearance, if any abnormality should immediately stop using and contact professionals. • Non-professionals should not disassemble or repair the equipment by themselves.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.