Overview
HUITAI DHP-9162N thermostatic incubator is a professional equipment for Universities & colleges, biology, agriculture, scientific research and other departments to store strains, biological foster and scientific research experiments. The product has a precise temperature control system and uniform temperature distribution to ensure the stability of the experimental environment.
Features
1. Large-screen LCD display, one-screen display of multiple sets of data, menu-type operation interface, simple to understand and easy to operate (standard)
2. Double-decked door structure, good thermal insulation performance, does not affect the temperature inside the box when observing the situation inside the box
3. Mirror stainless steel liner, diagonal air duct structure in the box, effectively ensure the Temperature Uniformity in the box, four corners and a half arc design is easy to clean, and the shelf spacing in the box is adjustable
4. The overall surface of the box sand blasting treatment, not easy to scratch, easy to clean
5. Electric heating film heating method, fast heating speed
6. The instrument control of the whole machine is placed in the upper left corner of the door panel, and the whole machine is beautiful and generous, saving product placement space
7. Equipped with 4 mobile casters, it is convenient for customers to move products
8. Independent temperature limit alarm system, which automatically interrupts when the temperature exceeds the limit to ensure the safe operation of the experiment without accidents (optional)
9. With RS485 interface, the recorder and computer can be connected, and the change of temperature parameters can be recorded (optional).
Principle
The electric thermostatic incubator quickly heats up through electric heating film heating, and adopts a precision instrument control system to realize temperature regulation. The diagonal air duct structure in the box ensures uniform circulation of hot air and maintains the temperature in the box within the set range to meet the needs of biological fostering and experiment.
Applications
Universities & colleges, biology, agriculture, research departments
Steps
1. Connect power supply AC220V 50HZ
2. Set the temperature parameters through the large-screen LCD display interface
3. Adjust the shelf spacing to meet the experimental needs
4. Start the equipment for heating and thermostatic control
5. Optional RS485 interface to connect the recorder or computer to monitor temperature changes
Notice
• The performance parameter test is carried out under no-load conditions
• When using, ensure that there is no strong magnetic and no vibration environment
• Recommended environmental conditions are temperature 30 ℃, humidity 50% RH
• Independent temperature limit alarm system is an optional function and requires additional configuration
• Use casters carefully when moving to avoid collisions