Overview
HUITAI HGZ-800N light chamber is a device used to simulate different environment climate conditions, which is widely used in the fields of microbial tissue cell fostering, seed germination, seedling experiment, plant cultivation, and insect and small animal feeding.
Features
1. Large-screen LCD display, one-screen display of multiple sets of data, menu-style operation interface, simple to understand and easy to operate.
2. Use 134a without fluorine refrigerant, energy saving and environmental protection.
3. Mirror stainless steel liner, four corners and a half arc design is easy to clean, and the shelf spacing in the box is adjustable.
4. The wind speed of the fan is adjusted by speed and speed to prevent the sample from losing moisture too quickly.
5. The instrument control of the whole machine is placed in the upper left corner of the door panel to save space.
6. There is a 25mm diameter test hole on the left side, which is convenient for experimental operation and temperature measurement.
7. Double-decked door structure, good thermal insulation performance, observation box does not affect the temperature.
8. The surface of the box is sand blasting, which is not easy to scratch and easy to clean.
9. Adopt brand Compressor and self-developed cooling system to prolong Compressor life.
10. Use JAKEL pipe flow circulating fan and unique air duct structure to ensure temperature uniformity in the box.
11. Standard LED light source, can be divided into 10 levels of dimming, Illuminant direction vertical, removable Illuminant board.
12. The microcomputer controls the temperature and illumination, and the temperature can be set differently day and night depending on the lighting.
13. PID control mode, accurate control of small fluctuations.
14. With 99 cycles, each cycle is divided into 30 segments, each of which is 99 hours and 59 minutes.
15. Supports long-term continuous operation.
16. Optional independent temperature limit alarm system, exceeding the limit temperature will automatically interrupt and sound and light alarm.
17. Optional USB or RS485 interface can be used to connect a printer or computer to record temperature parameters.
Principle
The light chamber precisely adjusts the temperature and illuminance through the microcomputer control system, and adopts PID control mode to ensure that the temperature fluctuation is small. The equipment uses brand Compressor and 134a without fluorine refrigerant for cooling, combined with JAKEL tube flow circulating fan and unique air duct structure to form a good air circulation system to ensure the temperature uniformity in the box. LED Illuminant can be divided into 10 levels of dimming, and can achieve different temperature settings day and night according to experimental needs.
Applications
Microbial tissue cell fostering, seed germination, seedling experiment, plant cultivation, insect, small animal feeding
Steps
1. Turn on the power supply and set the temperature and illumination parameters.
2. Adjust the fan wind speed gear as required.
3. Use the menu-style operation interface to program settings, and support multi-cycle cycles.
4. Optional connection to a printer or computer to record data.
5. Turn off the equipment after the experiment.
Notice
• Ambient Temperature should be in the range of 5~ 30 ℃.
• Power frequency AC380V 50Hz before use.
• Avoid opening the door frequently when the equipment is running, so as not to affect the temperature stability in the box.
• Regularly clean the liner and external surfaces to maintain equipment hygiene.
• Optional independent temperature limit alarm system to enhance experimental safety.