Overview
The QNix ® 9500 Probe is a new design, with a small and sturdy appearance, and is available in Fe, NFe and Fe/NFe dual-purpose models. The Probe has built-in LED lights to indicate whether the measurement value is over the limit, without the need to check the host display. The product is designed with an integrated built-in Probe, and an interchangeable external Probe with a Modular design is also available.
Features
1. Stainless steel Probe with protective film inside, IP65 dustproof and waterproof
2. Ruby Probe
3. Chemical resistant material, suitable for chemical cleaning (acetone, nitro diluted, ethanol)
4. Probe with LED indicator light
5. Probe line is drag chain type, can be bent arbitrarily
6. Each Probe memory electronic version calibrated certificate can be loaded into the test report through the PC
Principle
Using eddy Current measurement principle, the measurement range is 0~ 3000μm, and the error range is +/- 2~ 3%.
Applications
Coating thickness measurement, industrial testing, quality control
Standards
DIN EN ISO 2808、DIN 50981、DIN 50984、ISO 2178、BS 5411 (3 & 11)、BS 3900-C5、ASTM B 499、ISO 2360、ASTM D 1400、ASTM D 1186、ASTM D 7091
Steps
1. Connect the Probe to the host
2. Probe placed on the test surface for measurement
3. Observe the LED light prompt measurement status and over-limit situation
4. Load the electronic calibrated certificate of Probe memory to the test report through the PC
Notice
• Minimum measurement surface is φ 20mm
• Minimum radius of curvature is 20mm
• The thinnest substrate is 0.05mm.
• operating temperature -20~ 70 ℃
• Maximum workpiece surface temperature 80 ° C for continuous measurement
• Storage temperature is -20~ 70 ℃
• Probe clean to avoid affecting measurement accuracy