Overview
BENTING CBGDW-800L low/high temperature test chamber is a kind of equipment used for high and low temperature, constant heat and humidity reliability test of industrial products. The equipment has a wide temperature control range, and the performance index reaches the national standard GB10592-89, which is suitable for product testing in the fields of electronics and electricians, automobiles and motorcycles, aerospace and so on.
Features
1. The outer box is made of cold plate paint, and the inner wall is a good mirror stainless steel plate
2. The insulation material of the box is fiberglass
3. Equipped with single-opening hinged door, observation window, lighting and window frame/door frame anti-condensation electric heating device
4. The control panel includes a controller display screen, a U disk interface, an RS232 communication interface, and a sound and light alarm device
5. The unit includes a complete system such as a refrigeration unit, a water receiving tray, a drain hole, and a condensing fan
6. The power distribution system includes a variety of electrical components such as IO boards, machine tool transformers, ballasts, etc
7. The control system can automatically adjust the operation of the refrigerator to save energy
8. The evaporator cooling capacity is controlled by the solenoid valve switch
Principle
The test chamber drives the solenoid valve to switch the cooling capacity of the evaporator through the control system to achieve precise temperature control. The refrigeration system adopts a fully enclosed low-noise rotor Compressor and an air-cooled heat exchanger, and the working mode is -40 ° C~ -70 ° C. The binary cascade refrigeration method is used to ensure a stable temperature environment.
Applications
Electronics and electricians, automobiles and motorcycles, aerospace, ships and weapons, colleges and universities, R & D institutions
Standards
GB10592-89、GB2423.1、GB2423.2、GB2423.3、GJB150.3、GJB150.4、IEC、MIL
Steps
1. Connect the power supply and turn on the power switch
2. Set the required temperature parameters through the control panel
3. Start the device running program
4. Monitor the test process through the observation window
5. After the test is completed, turn off the device and disconnect the power supply
Notice
• Prohibit testing of products that may produce flammable, explosive, volatile, highly toxic, corrosive, and radioactive substances
• Make sure the equipment is well grounded before use
• Regularly inspect the refrigeration system and electrical components
• Keep equipment well ventilated
• Non-professionals should not disassemble or repair the equipment