Vertical airflow design prevents external backflow, wind speed adjustment range 0.2-0 m/sec, noise ≤ 62dB, working surface load-bearing 50kg, support multi-stage wind speed adjustment and sliding door arbitrary stop.
Category : Laminar Flow Cabinet
Brand : TaiSiTe
SKU : NB029138
Service :The delivery and service By NBCLab.Com
TAISITE CJ-2S Laminar Flow Cabinet is a double person vertical purification equipment with a flow direction from top to bottom, providing a 100-level clean environment. It is equipped with stainless steel interior and Cold rolled steel plate exterior, and the surface is drug-resistant painting, suitable for high cleanliness requirements of the work space.
1. Low noise design, touch button operation, support multi-level wind speed adjustment 2. Vertical air supply mode, sliding adjustment door can be parked arbitrarily 3. F-free series with integrated structure and LED indicator controller 4. -F series is a detachable split structure with LED indicator controller 5. -FS series is a detachable split structure with LCD intelligent controller 6. The split structure can be placed separately on the platform for use 7. Adopt Reinforced Glass and Quasi-Closed Table to Effectively Prevent External Air Flow
The Laminar Flow Cabinet draws air through an overhead fan, purifies it with a primary filter and a HEPA high-efficiency filter, and forms a vertical downward clean airflow. This one-way flow design effectively removes particles and microorganisms from the work area, maintaining 100 levels of cleanliness.
Biopharmaceuticals, electronics manufacturing, precision instruments, laboratory research, medical testing
Cleanliness level 100, 0.5μm particle collection efficiency above 99.9%, colony number ≤ 0.5/dish
1. Connect the AC220V power supply 2. Set working parameters through LED controller 3, adjust the wind speed to the required range (0.2-0 m/sec) 4. Turn on the UV Lamp for space disinfection 5. Start the fan to start the purification operation 6. Material handling through sliding adjustment doors
• The load-bearing capacity of the working surface shall not exceed 50kg • Regularly check Filter status and replace it in time • Keep the work area clean and avoid contamination • Make sure the horizontal adjustment foot is leveled before use • Pay attention to the safe use of UV Lamps and avoid direct exposure • Keep the environment ventilated when the equipment is running
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.