Overview
NELE GDH-3010 is an ultra-high accuracy Low Temperature Circulating Bath, Temperature range -30~ 100 ℃, Tank volume 10L, using fully enclosed environmentally friendly Compressor refrigeration system, built-in integrated new generation PID automatic temperature control program and High Accuracy PT100 temperature sensor to ensure accurate and stable temperature control.
Features
1. Fully enclosed environmentally friendly Compressor refrigeration system, using imported brands France Taikang or Japan Panasonic
2. Temperature range -30~ 100 ℃, the medium can be selected from pure water, ethylene glycol alcohol
3. Integrated new generation PID automatic temperature control program, built-in High Accuracy PT100 temperature sensor AAA level
4. Large screen highlight LCD screen, Chinese and English display, digital display resolution 0.0001 ℃
5. Adopt a non-heat centrifugal circulation pump to avoid the traditional pump's own heat generation affecting the temperature field of the tank body
6. The domestic circulation system is designed on the principle of porous U-shaped reflux circulation to achieve turbulence and high uniformity in the flow field and temperature field in the tank
7. The fuselage shell adopts anti-aging electrostatic spray technology, and the liner and countertop are made of S304 brushed stainless steel
8. Fine workmanship chassis front panel and countertop, unique user-friendly R rounded corner design structure
9. Multiple Safety protection devices: power-off protection, upper and lower limit temperature alarm, low water level protection and alarm, temperature loss protection and alarm, low water level anti-dry burning and explosion-proof heating device, Compressor overheating and overcurrent multiple protection
Principle
The equipment adopts a fully enclosed environmentally friendly Compressor refrigeration system and an integrated new generation of PID automatic temperature control program. The temperature is monitored in real time through the High Accuracy PT100 temperature sensor, and the non-heat centrifugal circulation pump and the porous U-shaped return circulation system are used to achieve turbulence and highly uniform distribution of the flow field and temperature field in the tank.
Applications
Laboratory temperature control, scientific research, industrial testing, mass testing
Standards
No relevant information yet
Steps
1. Select the appropriate medium (pure water, ethylene glycol alcohol) according to the experimental requirements.
2. Set the required temperature parameters
3. Power on the device
4. Monitor temperature changes through a large LCD screen
5. Power off the device after the experiment is completed
Notice
Make sure the equipment is in a horizontal position before use
• Regularly check the water level to avoid low water level operation
• Pay attention to well ventilated around the equipment
Do not move or tilt the device while it is running
• Regularly clean the inner and outer surfaces of the equipment
• In case of abnormal alarm, stop using it immediately and contact a professional