Overview
The RTS-60A low temperature verification tank is a NELE brand High Accuracy automatic temperature verification device, which has the characteristics of good temperature Stability, uniform temperature field and high temperature control accuracy. It adopts imported fully enclosed Compressor refrigeration system and intelligent PID adjustment to achieve automatic temperature control. The shell adopts all stainless steel structure and can be connected with RS-232 interface for users to connect the end point to realize computer control.
Features
1. High Accuracy Temperature Control, Temperature Fluctuation ≤ +/- 0.01 ℃
2. Good Temperature Field Uniformity, Temperature Uniformity ≤ 0.01 ℃
3. Adopt imported fully enclosed Compressor refrigeration system
4. Intelligent PID adjustment realizes automatic temperature control
5. With RS-232 interface, support computer control
6. All stainless steel shell structure
7. Temperature range: -60~ 100 ℃
Principle
The product achieves cooling through imported fully enclosed Compressor refrigeration system, and adopts intelligent PID adjustment technology to automatically control the temperature to ensure temperature Stability and uniformity. Users can connect to a computer through RS-232 interface for remote control.
Applications
Measurement, biochemical, petroleum, meteorology, energy, environmental protection, thermometer production, temperature controller production, temperature sensor production
Steps
1. Select the medium according to the temperature requirements: use absoluteethanol at low temperature, and use soft water at high temperature (above 30 ° C)
2. Connection power supply: 220V 50HZ
3. Set the target temperature through the panel or RS-232 interface
4. Start the device and wait for the temperature to stabilize
5. Perform temperature verification or experimental operation
Notice
• At low temperatures, absoluteethanol must be used as the medium
• At high temperatures (above 30 ° C), soft water is required as the medium
• Device weight up to 150kg, pay attention to safety when handling
• Power is 3kw, ensure that the power supply meets the requirements
• It takes about 120 minutes to raise and cool down, please arrange the experiment time reasonably