Overview
YH1000 haze meter (economical) can easily realize ASTM D1003 non-compensation method, full light transmitter, haze test, open sample compartment can be vertical and horizontal test, to adapt to more samples under test. The haze meter adopts PD array detector to meet the CIE V (lambda) 2 degree visual Response, and the compensation method can realize High Accuracy and repeability of light transmitter and haze data measurement.
Features
1. Conform to ASTM standards
2. Easy to operate, faster and more accurate measurement
3. Dynamic measurement
4. Easy measurement, wide sample adaptation
5. Quality management software
Principle
The YH1000 haze meter adopts transmission 0/D (parallel light illumination, diffuse reflection reception) illumination mode, uses 400~ 700nm combination LED Illuminant and PD array detector to meet the CIE V (lambda) 2 degree visual Response, and realizes the precise measurement of light transmitance and haze through the integrating sphere.
Applications
Glass Processing, Plastic Processing, Film, Display Processing, Encasement Industry, Liquid Liquid Analysis
Standards
ASTM D1003/1044、GB/T 2410、JJF 1303-2011、CIE 15.2、JIS K7105、JIS K7361、JIS K 7136
Steps
1. Instrument preheating correction
2. Place the sample in the open measurement area
3. Choose vertical or horizontal Test method
4. Large Touchscreen Operation
5. About 1.5 seconds to complete the measurement
6. Output data through USB interface
Notice
• Temperature range from 0 to 40 ° C (32 to 104 ° F)
• Storage Temperature range -20~ 50 ° C (-4~ 122 ° F)
• Sample thickness should be less than 170mm
• Power supply with DC 24V, 3A Power Adapter
• Ensure that the instrument preheating correction is completed before measurement