Overview
YOUKE YK3651B automatic Melt index meter is used to measure the melt flow rate of polymers in the viscous flow state. It is suitable for high temperature engineering plastics such as polycarbonic acid ester, polyaryl sulfone, fluoroplastics, and materials with low melting temperature such as polyethylene (PE), polystyrene, polypropylene (PP), ABS Resin, polyformaldehyde Resin, etc. The instrument is widely used in plastics production, plastic products, Petrochemicals and other industries, as well as related universities, R & D institutions and commodity inspection departments.
Features
1. Temperature range, up to 50~ 450 ℃.
2. Temperature control accuracy is high, Indication Error ≤ +/- 0.2 ℃, Fluctuation +/- 0.1 ℃.
3. Display resolution up to 0.1 seconds.
The temperature recovery time is short, ≤ 4 minutes.
5. Piston Displacement Indication Error is small, +/- 0.01mm.
Principle
The melt flow rate refers to the mass (MFR) or volume (MVR) of the thermal plasticity plastic through the standard die every 10 minutes under a certain temperature and load. This value can characterize the viscous flow characteristics of the thermal plasticity plastic in the molten state. The instrument tests according to this principle.
Applications
Plastic production, plastic products, Petrochemicals, related universities, R & D institutions, commodity inspection departments
Standards
GB/T 3682-2000、ISO 1133:2011、ASTM D 1238-2010
Steps
Please read the product instruction manual carefully for operation steps.
Notice
• Accessories and instruments should be placed in a drying environment.
• After the test, Accessories need to be coated with anti-rust grease, and the anti-rust grease needs to be removed before the test.
• After the test is completed, turn off the power supply to prevent accidents.