Overview
HONGKE GHP-300 intelligent light chamber is a 300L thermostatic light experiment equipment. It is suitable for cell tissue growth and cultivation, seed germination, seedling breeding, insect and small animal feeding, microbial antimicrobial foster preservation, especially for bioengineering, medical research, agricultural science, aquatic products, animal husbandry and other fields engaged in production and scientific research for thermostatic, light experiment foster.
Features
1. Using microcomputer technology, programmable control, temperature, Illumrination degree and time parameters can be set to simulate the day and night cycle lighting environment.
2. Touch switch is easy to operate, PID output control, temperature and Illumrination degree can be set separately during the day and night.
3. Illumrination degree 0~ 12000LX multi-level adjustable, with power-off memory function.
4. Intelligent thermostatic Control system Response fast, high temperature control accuracy, Temperature resolution of 0.1 ℃.
5. Air duct type ventilating, the air speed of the studio is soft, and the temperature is uniform (uniformity +/- 1 ℃).
6. With ultra-temperature Sensor abnormal protection function, and an independent air duct over-temperature protection device.
7. The four corners of the inner tank of the box are semi-circular arcs, using imported mirror or brushed stainless steel, with excellent manufacturing process and good thermal insulation performance.
Principle
Control system is used to program temperature, Illumrination degree and time. PID control algorithm is used to output control signals to precisely adjust heating, cooling and lighting systems to achieve a thermostatic and adjustable lighting environment. Ventilation is used to ensure uniform temperature inside the studio and simulate the day and night cycle of natural light, providing ideal conditions for biological growth or preservation.
Applications
Bioengineering, medical research, agricultural sciences, aquatic products, animal husbandry
Steps
Connect the 220V 50Hz power supply.
2. Perform parameter settings through Touchscreen, including temperature (10-50 ° C when there is light, 5-52 ° C when there is no light), Illumrination degree (0-12000LX adjustable) and Timer Range (1-9999 minutes).
3. Temperature and Illumrination degree of day and night can be set separately.
4. Place the sample on the shelf and close the box door.
5. Start the device and start running the program.
Notice
* Equipment Ambient Temperature must be between 5~ 35 ℃.
* Power supply voltage is 220V 50Hz.
Please confirm that the parameters are set correctly before running.
* Note that the maximum power consumption of the device is 950W.
* Do not use in environments that do not meet the specified conditions.
The equipment has a protective function, but it still requires regular inspection and maintenance.