Overview
JINGTIAN BY5 Standard Liquid is a Newtonian liquid used for inspection of Viscometer measurement results to determine whether Viscometer needs to be repaired or calibrated. Its Encasement specification is 500 ml/bottle, and the actual viscosity at 25 ° C is 4.688 mPa · s.
Features
1. Newton liquid characteristics to ensure the accuracy of inspection results.
Provide a clear actual viscosity value of 25 ° C (4.688mPa · s) as an inspection benchmark.
3. Standard Encasement Capacity (500ml/bottle) for easy use and storage.
Principle
As a Newtonian liquid of known viscosity, the standard solution has a stable viscosity value at a specific temperature (25 ° C). The measurement accuracy of the Viscometer can be inspected by comparing the measurement result with the known viscosity value of the standard solution using the Viscometer measurement.
Applications
Viscometer calibrating, Viscometer inspection, laboratory quality control
Standards
The data did not mention the specific standards met.
Steps
1. Make sure the Viscometer to be inspected is clean and drying.
2. Shake the BY5 standard solution well and stabilize the temperature at 25 ° C.
3. Use this standard solution for viscosity measurement according to Viscometer's operating procedures.
4. Record Viscometer measurement results.
5. Compare the measurement results with the actual viscosity value (4.688mPa · s) of the standard solution calibration.
6. Determine the measurement accuracy of the Viscometer based on the comparison results and decide whether it needs to be repaired or calibrated.
Notice
* Shake the standard solution well before use and ensure that the measurement temperature is stable at 25 ° C.
* Standard liquid should be stored in a cool, drying place, away from direct sunlight and high temperature.
* Please use it as soon as possible after opening and seal it to prevent contamination or volatilization.
* This product is a reference material for inspection and cannot be used for Miscellaneous purposes.
* The specific viscosity value is subject to the actual calibration value of the product.