$ 24892.00Ex works
Adopting multi-channel mechanical compression refrigeration system and frequency conversion technology to achieve 15 ° C/min linear temperature rise and drop; equipped with capillary tubing throttling system and forced air duct to ensure Temperature Uniformity ≤ 2 ° C; with remote monitoring and automatic fault diagnosis function.
Category : High Low Temperature Test Chamber
Brand : HUDA
SKU : NB041428
Service :The delivery and service By NBCLab.Com
HUDA QD-E708-512K40-X15 rapid temperature change test chamber is used to determine the adaptability of electronic products, automotive parts, materials, etc. under the conditions of high humidity, low humidity, constant heat and humidity, high and low temperature rapid alternating cycle, heat and humidity of surface condensation. The testing machine adopts reasonable structure and stable and reliable Control mode, with beautiful appearance, easy operation, safety, high temperature and humidity control accuracy. It is a professional equipment for rapid temperature change test, constant temperature and humidity test, alternating cabinet humidity test.
1. Adopt multi-channel, environmentally friendly and energy-saving (low noise) mechanical compression refrigeration system, use environmentally friendly materials and two-stage multi-channel main bypass system to achieve adaptive control and energy-saving effect, energy consumption only accounts for 80% of congeneric products, and use frequency conversion technology to reduce noise and vibration. 2. The self-developed multi-channel and multi-stage throttling system automatically adjusts the capillary tubing throttling through the intelligent control system to ensure adjustable flow and uniform temperature field. 3. The application of broadband compression equilibrate temperature control system, combined with temperature harmonizing chamber and centrifugal fan blades, to achieve uniform temperature distribution, and the PID adaptive algorithm developed in cooperation with South Korea Sanyuan Company to achieve fast convergence, small overshoot, Stability high temperature control. 4. Realize remote visual control, configure human-machine interface and RS485 communication, and support graphical operation, automatic fault prompt, and remote monitoring.
The test chamber intelligently controls the refrigeration output and flow rate through a multi-channel mechanical compression refrigeration system and a multi-stage throttling system to achieve rapid temperature change and uniform distribution. The broadband compression equilibrate temperature control system is adopted, combined with the temperature reconciliation chamber and centrifugal fan blades, to force air circulation to ensure full exchange of hot and cold and uniform air supply to the studio. The PID adaptive algorithm continuously corrects the control parameters to achieve rapid temperature convergence and stability.
Electronics, automotive parts, materials
GB2423.1-2008/IEC6008-2-1-2007、GB/T 2423.2-2008、GJB150.3A-2009、GJB150.4A-2009、GB2423.22-2008、GBT 2424.5-2006、GB/T 10589-2008、GB/T2423.3-2008(IEC68-2-3)、GJB150.9A-2009、G/BT 2423.4-2008/IEC6008-2-30:2005、GB/T5170.18-2005、GB/T10586-2006、GBT 2424.6-2006、GBT 2424.7-2006
1. Install the equipment in a flat and well-ventilated site to ensure that the load-bearing capacity is not less than 600kg/m ², and reserve maintenance space. 2. Connect the power supply (AC380V three-phase four-wire system + protective grounding) and check whether the voltage and frequency meet the requirements. 3. Set the temperature and humidity parameters through the touch screen controller, and select the program mode or the fixed value mode to operate. 4, Start equipment, monitor temperature and humidity changes, ensure uniformity and Stability. 5. Use the remote monitoring function to connect to the computer through the RS485 communication interface for real-time control. 6. After the test, turn off the equipment and check for trouble prompts. If necessary, follow the instructions to troubleshoot.
• The installation site must be free of flammable, explosive, corrosive gases and dust, and away from strong electromagnetic radiation sources. • Ensure that the ground load-bearing capacity is not less than 600kg/m ², and reserve enough maintenance space (A side is not less than 400mm, B side is not less than 400mm, C side is not less than 1000mm). • Power supply voltage allowable fluctuation range is 10%, frequency is 50 +/- 0.5Hz, protective ground resistance is less than 4Ω. • The water quality of the water supply must meet the second-level water standard of the National Analytical Laboratory, and the solid impurity content is less than 1mg/L. The working environment temperature should be 5~ 30 ℃, humidity ≤ 85% R.H. • When the equipment is running, pay attention to monitoring over-temperature, over-current and other protection devices, and deal with fault alarms in a timely manner. • Use the whole machine freight method during transportation to avoid violent vibration.
[A] No, it isn't. This is the ex-factory price of the product in the Chinese Mainland. In addition to this price, buyers from different countries need to account for additional costs such as shipping fees, tariffs, and others. For specific regional quotes, please consult our customer service or fill out a quotation form.
[A] First, please click on the quotation form on the website and fill in detailed information. Our customer service personnel will then send you a quotation based on your specific country and region, confirming payment and delivery methods. Following this, we proceed with contract signing, payment, and shipment. You can then await delivery.
[A] After-sales repair for overseas instruments is an unavoidable issue. Our repair terms stipulate that the sender is responsible for the shipping and insurance costs incurred for the repair. Repairs within the warranty period are free, while repair costs beyond the warranty period are determined based on the actual situation.