Overview
TR 1202/050 double-channel Fineness Gauge is a TRUIT brand double-channel Grindometer made of high-quality wear-resistant and corrosion-resistant alloy steel with a Hardness of 52 +/- 2HRC to ensure a long service life. With a range of 0~ 50μm, Graduation 2.5μm, groove size 140 × 12.5mm, it is suitable for measurement of a variety of materials.
Features
1. Made of high-quality wear-resistant and corrosion-resistant alloy steel, Hardness 52 +/- 2HRC ensures service life
2. double-channel design with 2 grooves
3. Range 0~ 50μm, Graduation 2.5μm
4. Groove size 140 × 12.5mm
5. Suitable for a variety of material measurement, widely used
Principle
Grindometer scraping the sample on the trench by a scraper, measuring the size and dispersion of the pigment and impurity particles according to the distribution of the sample in the trench, thereby controlling the mass of the dispersed product in production, storage and application.
Applications
Coatings, lacquers, plastics, inks, paper, Pottery and porcelain, food, cosmetics, and other liquids and pastes
Standards
GB/T 1724、GB/T 6753.1、GB/T 13217.3、ISO 1524
Steps
1. Prepare the sample to be tested
2. Place the sample on the Fineness Gauge
3. Scrape evenly across the grooves with a scraper
4. Observe the distribution of the sample in the trench
5. Read particle size and dispersion data
Notice
• Fineness Gauge drying before use
• The scraper should be operated with uniform force
• Avoid Hard Objects Scratching Fineness Gauging Surfaces
• Clean and maintain in time after use
• Avoid wet environments when storing