Overview
FRARIV C200N smart tap density meter is a precision testing instrument suitable for metal powder, Non-metallic powder, chemical industry, new energy, New material and scientific research and teaching. The instrument adopts advanced technology, has High Accuracy measurement and convenient operation characteristics, and can automatically read powder sample mass and generate electronic reports.
Features
1.4.3 inch touch screen, providing convenient, simple and natural human-machine interaction
2. Long Life, High Reliability Motor Components with Start-Stop Response
3. Can automatically read the powder sample mass from the electronic scale, improve the convenience of operation
4. Adopt 32-bit embedded RISC microprocessor, featuring low cost, high performance, and low power consumption
5. Use Customized Embedded Operating System with Strong System Stability and Software Expansion Capability
6. Adopt lightweight, high-performance, highly reliable, configurable GUI interface to enhance user experience
7. Built-in embedded micro-receipt printer
8. Calculation results are accurate to 3 decimal places
9. Have good network connection ability, support remote software upgrade and customization needs
10. Support electronic report generation and download, providing Excel and PDF formats
11. The instrument has a solid structure, beautiful appearance and easy operation
Principle
The vibration density meter processes powder samples with a vibration frequency of 250 rpm and a vibration amplitude of 3 mm by swirl/spin. The instrument automatically reads the powder sample mass from the electronic scale, and calculates it accurately through the built-in 32-bit embedded RISC microprocessor. Finally, the vibration density value is obtained and the test report is generated.
Applications
Metal Powder, Non-metallic Powder, Chemical Industry, New Energy, New material, Research and Teaching
Standards
Chinese Pharmacopoeia 2020 Edition
Steps
1. Preparation of powder samples and measuringflasks
2. Set test parameters through the touch screen
3. The instrument automatically reads the powder mass from the electronic scale
4. Start the vibration program for testing
Step 5 Review the calculation results
6. Use the built-in printer to print reports or generate electronic reports
Notice
• Ensure that the working power supply is AC220V 50Hz
• Check whether the measuringflask volume meets the 250mL requirement before use
• Note that repeatability errors should be controlled within the range of ≤ 1%
• Built-in printer paper width is 58mm, paper compartment capacity ≤ 30mm
• The roll length is 9.42m and the printing speed is 60mm/s.
• Regularly check the structural integrity of the instrument
• Follow relevant safety regulations when operating