Overview
The MOLECULAR MP-40L reverse osmosis (RO) water purifier is a device used for laboratory physical and chemical quantitative analysis, buffer solution and high standard solution preparation. It includes PP fiber filter element, activated carbon filtrate, reverse osmosis (RO) membrane, purification column and other accessories. The water production rate of the equipment is 20L/h, the water intake Flow velocity is 1.5L/min, and it is controlled by intelligent microcomputer. It has the functions of operation status display and water quality online monitoring.
Features
1. Fully automatic microcomputer control to achieve automatic water generation, standby, alarm, flushing and other functions
2. RO film intelligent anti-scale regular rinsing and manual forced rinsing
3. Automatic domestic circulation system to maintain ultra-pure water quality
4. Equipped with PE controllable 30-liter pure water tank
5. Using high-quality components including DOW membrane, ROHMHAAS ultra-pure Resin, MOL filter, etc
6. Online display of water resistivity, LCD display interface
7. With automatic protection functions such as water failure, power failure, and overload
Principle
The equipment purifies water through a multi-stage filtrate system such as PP fiber filter element, activated carbon filtrate, reverse osmosis (RO) membrane and purification column. The special ion exchange Resin and special flow path design for ultra-pure water in the United States are used to effectively remove trace ionic substances in water and achieve efficient purification.
Applications
Laboratory physical and chemical quantitative analysis, buffer solution and high standard solution preparation
Standards
GB ⅲ pure water standard, GB6682 ⅰ water standard
Steps
1. Connect the tap water source to ensure that the water pressure 0.1-0
2. Connect the 220V/50HZ power supply
3. Start the device by touching the button
4. Equipment automatically generates water and displays operating status
5. Monitoring online water quality parameters
6. Use pure water from the PE pure water tank
Notice
• The working environment water temperature should be kept at 5~ 40 ℃
• The northern region needs to be used in an antifreeze room
• Regularly check the status of filter elements and membrane modules
• Pay attention to the equipment alarm prompt and deal with it in time
• Ensure that the influent quality meets the requirements